气相色谱-质谱%氢气%载气%自动质谱解卷积鉴定系统%解良度%有机磷农药
氣相色譜-質譜%氫氣%載氣%自動質譜解捲積鑒定繫統%解良度%有機燐農藥
기상색보-질보%경기%재기%자동질보해권적감정계통%해량도%유궤린농약
gas chromatography-mass spectrometry( GC-MS)%hydrogen%carrier gas%Auto-mated Mass Spectral Deconvolution & Identification System( AMDIS)%fraction good%organo-phosphorus pesticides
迄今为止,氦气几乎是气相色谱-质谱联用仪运行中唯一可选的载气。本文使用气相色谱-质谱联用仪考察了氢气和氦气分别作为载气时,对5种有机磷农药混合标准溶液的检测,以确定氢气代替氦气作为气相色谱-质谱联用仪载气用于有机磷农药分析的可行性。用工作站自带解卷积软件(自动质谱解卷积鉴定系统,AMDIS32),结合自建“有机农药质谱库”对结果进行了解析,而后用 SPSS 统计软件对数据进行了聚类分析。结果发现,以氢气为载气时虽然出峰稍快,但色谱峰分离度较低,解良度较低,且待测物信号不稳定,测定准确度较低,有些化合物甚至检测不到。因此以氢气为载气分析有机磷农药时应对测定准确度、信号稳定性、安全性等诸多因素进行慎重考虑。
迄今為止,氦氣幾乎是氣相色譜-質譜聯用儀運行中唯一可選的載氣。本文使用氣相色譜-質譜聯用儀攷察瞭氫氣和氦氣分彆作為載氣時,對5種有機燐農藥混閤標準溶液的檢測,以確定氫氣代替氦氣作為氣相色譜-質譜聯用儀載氣用于有機燐農藥分析的可行性。用工作站自帶解捲積軟件(自動質譜解捲積鑒定繫統,AMDIS32),結閤自建“有機農藥質譜庫”對結果進行瞭解析,而後用 SPSS 統計軟件對數據進行瞭聚類分析。結果髮現,以氫氣為載氣時雖然齣峰稍快,但色譜峰分離度較低,解良度較低,且待測物信號不穩定,測定準確度較低,有些化閤物甚至檢測不到。因此以氫氣為載氣分析有機燐農藥時應對測定準確度、信號穩定性、安全性等諸多因素進行慎重攷慮。
흘금위지,양기궤호시기상색보-질보련용의운행중유일가선적재기。본문사용기상색보-질보련용의고찰료경기화양기분별작위재기시,대5충유궤린농약혼합표준용액적검측,이학정경기대체양기작위기상색보-질보련용의재기용우유궤린농약분석적가행성。용공작참자대해권적연건(자동질보해권적감정계통,AMDIS32),결합자건“유궤농약질보고”대결과진행료해석,이후용 SPSS 통계연건대수거진행료취류분석。결과발현,이경기위재기시수연출봉초쾌,단색보봉분리도교저,해량도교저,차대측물신호불은정,측정준학도교저,유사화합물심지검측불도。인차이경기위재기분석유궤린농약시응대측정준학도、신호은정성、안전성등제다인소진행신중고필。
Helium is almost the only choosable carrier gas used in gas chromatography-mass spectrometry( GC-MS ). A mixed standard solution of five organophosphorus pesticides was analyzed by using GC-MS,and hydrogen or helium as carrier gas,so as to study the feasibility of hydrogen instead of helium as carrier gas for the determination of organophosphorus pesti-cides. Combining a mass spectrum database built by ourselves,the results were deconvolved and identified by Automated Mass Spectral Deconvolution &Identification System( AMDIS32), a software belonging to the workstation of the instrument. Then,the statistical software,IBM SPSS Statistics 19. 0 was used for the clustering analysis of the data. The results indicated that when hydrogen was used as carrier gas,the peaks of the pesticides detected were slightly earli-er than those when helium used as carrier gas,but the resolutions of the chromatographic peaks were lower,and the fraction good indices( Frac. Good)were lower,too. When hydro-gen was used as carrier gas,the signals of the pesticides were unstable,the measuring accura-cies of the pesticides were reduced too,and even more,some compounds were undetectable. Therefore,considering the measuring accuracy, the signal stability, and the safety, etc., hydrogen should be cautiously used as carrier gas in the determination of organophosphorus pesticides by GC-MS.