CHINA MEDICAL EQUIPMENT
2014年
12期
112-114,118
,共4页
刘海%杭霞瑜%李益坤%张新良%丁巍
劉海%杭霞瑜%李益坤%張新良%丁巍
류해%항하유%리익곤%장신량%정외
螺旋断层放射治疗%剂量验证%γ分析%胶片数字化仪%质量保证
螺鏇斷層放射治療%劑量驗證%γ分析%膠片數字化儀%質量保證
라선단층방사치료%제량험증%γ분석%효편수자화의%질량보증
helical tomotherapy%dosimetric verification%γ analysis%film digitizer%quality assurance
目的:探讨OmniPro I’mRT分析软件(OI)与平板式RGB扫描仪(FS)在螺旋断层放疗计划验证中的应用价值。方法随机抽取12例螺旋断层放疗计划,将GAFchromic EBT3自显影胶片置于CheesePhantom模体中进行照射以获取冠状面剂量分布,采用γ分析法对OI+FS方案、螺旋断层放疗计划(TomoPS)DQA分析软件+Vidar胶片数字化仪(VidarFD)方案进行比较分析。结果 TomoPS +VidarFD方案的平均通过率为(97.32%±1.16%),OI+FS方案的平均通过率为(96.40%±1.01%),两者通过率均较高;Pearson相关性分析显示两组数据具有极强相关性(r=0.919,P<0.05)。结论OmniPro I’mRT分析软件配合平板RGB扫描仪可作为螺旋断层放疗计划验证的手段。
目的:探討OmniPro I’mRT分析軟件(OI)與平闆式RGB掃描儀(FS)在螺鏇斷層放療計劃驗證中的應用價值。方法隨機抽取12例螺鏇斷層放療計劃,將GAFchromic EBT3自顯影膠片置于CheesePhantom模體中進行照射以穫取冠狀麵劑量分佈,採用γ分析法對OI+FS方案、螺鏇斷層放療計劃(TomoPS)DQA分析軟件+Vidar膠片數字化儀(VidarFD)方案進行比較分析。結果 TomoPS +VidarFD方案的平均通過率為(97.32%±1.16%),OI+FS方案的平均通過率為(96.40%±1.01%),兩者通過率均較高;Pearson相關性分析顯示兩組數據具有極彊相關性(r=0.919,P<0.05)。結論OmniPro I’mRT分析軟件配閤平闆RGB掃描儀可作為螺鏇斷層放療計劃驗證的手段。
목적:탐토OmniPro I’mRT분석연건(OI)여평판식RGB소묘의(FS)재라선단층방료계화험증중적응용개치。방법수궤추취12례라선단층방료계화,장GAFchromic EBT3자현영효편치우CheesePhantom모체중진행조사이획취관상면제량분포,채용γ분석법대OI+FS방안、라선단층방료계화(TomoPS)DQA분석연건+Vidar효편수자화의(VidarFD)방안진행비교분석。결과 TomoPS +VidarFD방안적평균통과솔위(97.32%±1.16%),OI+FS방안적평균통과솔위(96.40%±1.01%),량자통과솔균교고;Pearson상관성분석현시량조수거구유겁강상관성(r=0.919,P<0.05)。결론OmniPro I’mRT분석연건배합평판RGB소묘의가작위라선단층방료계화험증적수단。
Objective To investigate the application value of OmniPro I’mRT (OI)+RGB fiatbed scanner (FS) in the verification of helical tomotherapy (HT) plans.Methods 12 HT plans were randomly selected, and GAFchromic EBT3 films placed in CheesePhantom were irradiated to obtain the coronal plane dose distribution. Then the γanalysis was used in comparing and analyzing the verification effects of OI+FS program and the tomo planning station (TomoPS) DQA analysis software+ Vidar film digitizer (VidarFD) program.ResuItsThe average pass rate of TomoPS+VidarFD program was (97.32%±1.16%), while that of OI+FS program was (96.40%±1.01 %), which indicated that the average pass rates of the two programs were high. The Pearson correlation analysis results showed that the verification data of OI+FS program was closely related to that of TomoPS+VidarFD program (r=0.919,P<0.05). ConcIusionOmniPro I ‘mRT analysis software combined with RGB fiated scanner can be used as a method in verifying the HT plans.