有核红细胞%雅培CELL-DYN Ruby%显微镜镜检
有覈紅細胞%雅培CELL-DYN Ruby%顯微鏡鏡檢
유핵홍세포%아배CELL-DYN Ruby%현미경경검
Nucleated red blood cells%Abbott CELL-DYN Ruby%Microscopic examination
目的:探讨雅培CELL-DYN Ruby全自动血细胞分析仪检测有核红细胞(NRBC)的准确性。方法2012年3月—2014年2月取镜检确认的外周血涂片有NRBC的标本102例,分别用显微镜镜检计数法和细胞计数仪法两种方法计数,得到两组数据进行比较分析。结果当标本中NRBC百分比<5%时,两种检测方法比较,差异无统计学意义(P>0.05),说明雅培CELL-DYN Ruby全自动血细胞分析仪检测有核红细胞准确。当标本中NRBC百分比≥5%时,差异有统计学意义,两种检测方法有明显差异,说明在此种情况下检测有一定偏差。结论雅培CELL-DYN Ruby全自动血细胞分析仪检测有核红细胞时准确性较差,如在检测时有NRBC的提示时,一定要推血涂片镜检计数,并用公式计算出白细胞总数。
目的:探討雅培CELL-DYN Ruby全自動血細胞分析儀檢測有覈紅細胞(NRBC)的準確性。方法2012年3月—2014年2月取鏡檢確認的外週血塗片有NRBC的標本102例,分彆用顯微鏡鏡檢計數法和細胞計數儀法兩種方法計數,得到兩組數據進行比較分析。結果噹標本中NRBC百分比<5%時,兩種檢測方法比較,差異無統計學意義(P>0.05),說明雅培CELL-DYN Ruby全自動血細胞分析儀檢測有覈紅細胞準確。噹標本中NRBC百分比≥5%時,差異有統計學意義,兩種檢測方法有明顯差異,說明在此種情況下檢測有一定偏差。結論雅培CELL-DYN Ruby全自動血細胞分析儀檢測有覈紅細胞時準確性較差,如在檢測時有NRBC的提示時,一定要推血塗片鏡檢計數,併用公式計算齣白細胞總數。
목적:탐토아배CELL-DYN Ruby전자동혈세포분석의검측유핵홍세포(NRBC)적준학성。방법2012년3월—2014년2월취경검학인적외주혈도편유NRBC적표본102례,분별용현미경경검계수법화세포계수의법량충방법계수,득도량조수거진행비교분석。결과당표본중NRBC백분비<5%시,량충검측방법비교,차이무통계학의의(P>0.05),설명아배CELL-DYN Ruby전자동혈세포분석의검측유핵홍세포준학。당표본중NRBC백분비≥5%시,차이유통계학의의,량충검측방법유명현차이,설명재차충정황하검측유일정편차。결론아배CELL-DYN Ruby전자동혈세포분석의검측유핵홍세포시준학성교차,여재검측시유NRBC적제시시,일정요추혈도편경검계수,병용공식계산출백세포총수。
Objective To investigate the accuracy of Abbott CELL-DYN Ruby automatic blood cell analyzer for the detection of nucleated red blood cell (NRBC). Methods 102 peripheral blood smears with NRBC confirmed by microscopic examination were taken and counted by microscopy counting method and cell counter method, respectively. And the two sets of data were compared and analyzed. Results There was no statistically significant difference between the two test methods when the NRBC containing in the specimens accounted <5%, P>0.05, which suggested that Abbott CELL-DYN Ruby automatic blood cell analyzer could detect NRBC accurately. When the NRBC containing in the specimens accounted ≥5%, the difference between the two test methods was statistically significant, which showed that the detection had certain deviation under this kind circumstance. Conclusion Abbott CELL-DYN Ruby automatic blood cell analyzer has poor accuracy in detecting the nucleated red blood cells. If the detection suggests that there are nucleated red blood cells, the blood specimen should be smeared for microscopy counting, and the total white blood cells should be calculated by the formula.