中国医疗设备
中國醫療設備
중국의료설비
CHINA MEDICAL EQUIPMENT
2015年
4期
104-106
,共3页
潘国兵%孙建军%帅海涛%吴慧敏
潘國兵%孫建軍%帥海濤%吳慧敏
반국병%손건군%수해도%오혜민
容积CT%头颅扫描%噪声指数%自动毫安技术%辐射剂量
容積CT%頭顱掃描%譟聲指數%自動毫安技術%輻射劑量
용적CT%두로소묘%조성지수%자동호안기술%복사제량
volume computerized tomography%cranial computerized tomography scanning%noise indexes%automatic mA technology%radiation dose
目的:探讨噪声指数对64排容积CT头颅扫描图像质量及辐射剂量的影响。方法对我院收治的90例脑梗塞患者复查时行64排容积CT扫描,设置不同的噪声指数,通过主观和客观评价对比说明噪声指数对图像质量的影响。结果在非螺旋模式下,当噪声指数<4时,图像质量差异不明显;当噪声指数在4~6时,图像质量差异较为模糊;当噪声指数>6时,图像质量明显下降,不能满足诊断要求。结论应在满足临床诊断前提下合理设置容积CT扫描的噪声指数。
目的:探討譟聲指數對64排容積CT頭顱掃描圖像質量及輻射劑量的影響。方法對我院收治的90例腦梗塞患者複查時行64排容積CT掃描,設置不同的譟聲指數,通過主觀和客觀評價對比說明譟聲指數對圖像質量的影響。結果在非螺鏇模式下,噹譟聲指數<4時,圖像質量差異不明顯;噹譟聲指數在4~6時,圖像質量差異較為模糊;噹譟聲指數>6時,圖像質量明顯下降,不能滿足診斷要求。結論應在滿足臨床診斷前提下閤理設置容積CT掃描的譟聲指數。
목적:탐토조성지수대64배용적CT두로소묘도상질량급복사제량적영향。방법대아원수치적90례뇌경새환자복사시행64배용적CT소묘,설치불동적조성지수,통과주관화객관평개대비설명조성지수대도상질량적영향。결과재비라선모식하,당조성지수<4시,도상질량차이불명현;당조성지수재4~6시,도상질량차이교위모호;당조성지수>6시,도상질량명현하강,불능만족진단요구。결론응재만족림상진단전제하합리설치용적CT소묘적조성지수。
Objective To select the appropriate noise index of the GE 64-Row VCT (Volume Computerized Tomography) Scanner in cranial CT (Computerized Tomography) so as to reasonably reduce radiation dose while meeting the requirements of clinical diagnosis. Methods Different noise indexes of the 64-Row VCT scanner were set for the patients with cerebral infarction who underwent a re-examination of cranial CT in the hospital. Then, comparisons were made through subjective and objective evaluation of the quality of images acquired in different noise indexes. Results Under the non-spiral mode, no obvious differences were found in the quality of images when the noise index < 4; the quality of images had indistinct differences but satisifed the diagnostic requirements when the noise index was between 4 to 6;the image quality was obviously decreased and could not meet the diagnostic requirements when the noise index>6. Conclusion The noise index should be set individually in consideration of the patient radiation dose while the requirements of the clinical diagnosis were also satisifed.