Academic Journal of Guangzhou Medical College
2014年
6期
74-76
,共3页
黄奋尧%俞家熙%陈智杰%方鸣%宋玉全
黃奮堯%俞傢熙%陳智傑%方鳴%宋玉全
황강요%유가희%진지걸%방명%송옥전
胸部摄影%防护装置%辐射剂量
胸部攝影%防護裝置%輻射劑量
흉부섭영%방호장치%복사제량
chest radiography%protective device%radiation dose
目的::探讨胸部立位X线摄影防护装置的防护效果。方法:设计与探测板联动的防护装置用于防护颈部和腹部,通过比较防护装置使用前后的胸部正侧位摄影条件和剂量仪数据,评价该防护装置的防护效果。结果:实验组(使用防护装置,50例)正侧位摄影均值(1.6270±0.3098)mAs,(3.6298±1.4268) mAs;对照组(未使用防护装置,50例)正侧位摄影均值为(1.6580±0.3742)mAs,(3.6800±1.2521)mAs。实验组正侧位探测剂量均值为(376.1514±96.5325)nGy,(735.8842±272.8077)nGy;对照组正侧位探测剂量均值为(506.9260±180.5912)nGy,(1178.9160±535.2549)nGy。两组正侧位摄影条件差异无统计学意义( P>0.05),剂量差异有统计学意义( P<0.05)。结论:胸部立位X线摄影防护装置的使用可有效减低防护部位的辐射剂量。
目的::探討胸部立位X線攝影防護裝置的防護效果。方法:設計與探測闆聯動的防護裝置用于防護頸部和腹部,通過比較防護裝置使用前後的胸部正側位攝影條件和劑量儀數據,評價該防護裝置的防護效果。結果:實驗組(使用防護裝置,50例)正側位攝影均值(1.6270±0.3098)mAs,(3.6298±1.4268) mAs;對照組(未使用防護裝置,50例)正側位攝影均值為(1.6580±0.3742)mAs,(3.6800±1.2521)mAs。實驗組正側位探測劑量均值為(376.1514±96.5325)nGy,(735.8842±272.8077)nGy;對照組正側位探測劑量均值為(506.9260±180.5912)nGy,(1178.9160±535.2549)nGy。兩組正側位攝影條件差異無統計學意義( P>0.05),劑量差異有統計學意義( P<0.05)。結論:胸部立位X線攝影防護裝置的使用可有效減低防護部位的輻射劑量。
목적::탐토흉부립위X선섭영방호장치적방호효과。방법:설계여탐측판련동적방호장치용우방호경부화복부,통과비교방호장치사용전후적흉부정측위섭영조건화제량의수거,평개해방호장치적방호효과。결과:실험조(사용방호장치,50례)정측위섭영균치(1.6270±0.3098)mAs,(3.6298±1.4268) mAs;대조조(미사용방호장치,50례)정측위섭영균치위(1.6580±0.3742)mAs,(3.6800±1.2521)mAs。실험조정측위탐측제량균치위(376.1514±96.5325)nGy,(735.8842±272.8077)nGy;대조조정측위탐측제량균치위(506.9260±180.5912)nGy,(1178.9160±535.2549)nGy。량조정측위섭영조건차이무통계학의의( P>0.05),제량차이유통계학의의( P<0.05)。결론:흉부립위X선섭영방호장치적사용가유효감저방호부위적복사제량。
Objective:To investigate the effect of the protective device of orthostatic chest X ̄ray radiography.Methods:The board ̄linked protective device on neck and abdomen was designed and determined.The protective effect of the device was evaluated by comparing the radiographic condition and dosimeter data of the anterior and lateral positions in chest before and after use. Results: The mean mAs values of radiography in the anterior and lateral positions of the chest in the experimental group ( with protective device) and control group (without protective device) were(1.627 0±0.309 8)and(3.629 8±1.426 8)vs(1.658 0±0.374 2) and (3.680 0± 1. 252 1 ) , respectively. The mean value of the detection dose on the anterior and lateral positions in the experimental group and control group were (376.151 4±96.532 5) and (735.8842±272.8077) vs (506.926 0± 180.591 2 )vs (1 178.9160±535.254 9), respectively.There was no statistically significant difference in the radiographic condition between the experimental group and control group ( P>0. 05 ) . However, statistically significant differences were found in the dose between the two groups(P<0.05).Conclusion:Using the protective device of orthostatic chest X ̄ray radiography can effectively reduce the radiation dose in the protected positions.