目的 使用自主知识产权轴向磁力卸载微型植入式心脏轴流泵进行动物实验,研究轴流泵的血流动力学表现、心室去负荷效果和血液相容性.方法 实验用猪5只,体质量(40.0±5.2)kg.全麻后于左侧第5肋间开胸,植入辅助泵,泵入口引流管经心尖插入左心室,出口用人工血管与降主动脉行端侧吻合,留置左心室测压管.结果 辅助泵完全植入动物胸腔内,手术成功率100%.手术后5只实验猪均存活1周以上.辅助泵的转速与辅助效果呈正相关性.平均心室收缩压(131.6 ±28.0)mmHg(1 mmHg =0.133 kPa),辅助时左心室舒张末期压力降低,心室内平均压随之减低.外周血流动力学稳定,辅助前、后外周循环血压无明显差异.术后1周内实验猪尿量稳定.实验猪血浆中游离血红蛋白在手术日轻微升高,术后第1天至1周逐渐恢复至术前水平.华法林3 mg/天口服抗凝,INR值维持在2.0~2.5,标本检查未发现泵内血栓形成.结论 自主知识产权轴向磁力卸载微型植入式心脏轴流泵应用于实验猪的血流动力学稳定,心脏压力去负荷效果良好,能维持良好的外周循环灌注,且具良好的血液相容性.
目的 使用自主知識產權軸嚮磁力卸載微型植入式心髒軸流泵進行動物實驗,研究軸流泵的血流動力學錶現、心室去負荷效果和血液相容性.方法 實驗用豬5隻,體質量(40.0±5.2)kg.全痳後于左側第5肋間開胸,植入輔助泵,泵入口引流管經心尖插入左心室,齣口用人工血管與降主動脈行耑側吻閤,留置左心室測壓管.結果 輔助泵完全植入動物胸腔內,手術成功率100%.手術後5隻實驗豬均存活1週以上.輔助泵的轉速與輔助效果呈正相關性.平均心室收縮壓(131.6 ±28.0)mmHg(1 mmHg =0.133 kPa),輔助時左心室舒張末期壓力降低,心室內平均壓隨之減低.外週血流動力學穩定,輔助前、後外週循環血壓無明顯差異.術後1週內實驗豬尿量穩定.實驗豬血漿中遊離血紅蛋白在手術日輕微升高,術後第1天至1週逐漸恢複至術前水平.華法林3 mg/天口服抗凝,INR值維持在2.0~2.5,標本檢查未髮現泵內血栓形成.結論 自主知識產權軸嚮磁力卸載微型植入式心髒軸流泵應用于實驗豬的血流動力學穩定,心髒壓力去負荷效果良好,能維持良好的外週循環灌註,且具良好的血液相容性.
목적 사용자주지식산권축향자력사재미형식입식심장축류빙진행동물실험,연구축류빙적혈류동역학표현、심실거부하효과화혈액상용성.방법 실험용저5지,체질량(40.0±5.2)kg.전마후우좌측제5륵간개흉,식입보조빙,빙입구인류관경심첨삽입좌심실,출구용인공혈관여강주동맥행단측문합,류치좌심실측압관.결과 보조빙완전식입동물흉강내,수술성공솔100%.수술후5지실험저균존활1주이상.보조빙적전속여보조효과정정상관성.평균심실수축압(131.6 ±28.0)mmHg(1 mmHg =0.133 kPa),보조시좌심실서장말기압력강저,심실내평균압수지감저.외주혈류동역학은정,보조전、후외주순배혈압무명현차이.술후1주내실험저뇨량은정.실험저혈장중유리혈홍단백재수술일경미승고,술후제1천지1주축점회복지술전수평.화법림3 mg/천구복항응,INR치유지재2.0~2.5,표본검사미발현빙내혈전형성.결론 자주지식산권축향자력사재미형식입식심장축류빙응용우실험저적혈류동역학은정,심장압력거부하효과량호,능유지량호적외주순배관주,차구량호적혈액상용성.
Objective To explore the in-vivo hemodynamic and hemolysis effect of a newly designed axial continuousflow ventricular assist device(VAD) in swine.Methods Under general anesthesia,each of 5 swine [weight (40.0 ± 5.2)kg] was implanted with the axial continuous-flow VAD into the apex of left heart ventricle,and the outflow graft was anastomosised to descending aorta.Results All of the axial continuous-flow VAD were implanted successfully with post-operative survival rate 100%.All 5 animals survived over one week.There was a positive correlation between pump speed and assistance effect.The mean left ventricular systolic pressure was (131.6 ± 28.0) mmHg(1 mmHg =0.133 kPa).While the axial continuous-flow VAD was working,left ventricular end diastolic pressure decreased,along with mean intraventricular pressure declined.Peripheral hemodynamics was stable and peripheral blood pressure was not remarkably different from the pressure preoperation.Daily urine volume was in normal range within 1 week post operation.Free hemoglobin in plasma was slightly elevated on the surgery day,and gradually dropped to normal level within 1 week.International Normalized Ratio(INR) was maintained between 2.0-2.5 with oral adminiatration of warfarin of 3 mg/day.There was no thrombosis existing in VAD at autopsy.Conclusion The application of the axial pump with hydrodynamic-magnetically levitated impeller in animal experiment can provide stable hemodynamics,advanced heart unloaded effect,favorable peripheral perfusion,and blood compatibility is satisfactory.