研制新型的基于双通道脉搏传感的腕戴式无线低功耗心率实时监测装置.根据人体腕部生理解剖的特点,提出桡动脉、尺动脉双通道脉搏波及差分信号同步检测的新方法.研制了集传感器、调理电路、微处理器、通讯、自动增益算法和心率算法等软、硬件部件于一体的小型化装置.电路系统测试表明:在待机和工作模式下的平均工作电流分别约为10和300μA ;对10名男性在休憩状态下的心率进行3 h的连续动态监测,将测试结果与标准动态心电图记录的数据进行比较,结果表明,心率测量的算术平均误差约为0.3 BPM .
研製新型的基于雙通道脈搏傳感的腕戴式無線低功耗心率實時鑑測裝置.根據人體腕部生理解剖的特點,提齣橈動脈、呎動脈雙通道脈搏波及差分信號同步檢測的新方法.研製瞭集傳感器、調理電路、微處理器、通訊、自動增益算法和心率算法等軟、硬件部件于一體的小型化裝置.電路繫統測試錶明:在待機和工作模式下的平均工作電流分彆約為10和300μA ;對10名男性在休憩狀態下的心率進行3 h的連續動態鑑測,將測試結果與標準動態心電圖記錄的數據進行比較,結果錶明,心率測量的算術平均誤差約為0.3 BPM .
연제신형적기우쌍통도맥박전감적완대식무선저공모심솔실시감측장치.근거인체완부생리해부적특점,제출뇨동맥、척동맥쌍통도맥박파급차분신호동보검측적신방법.연제료집전감기、조리전로、미처리기、통신、자동증익산법화심솔산법등연、경건부건우일체적소형화장치.전로계통측시표명:재대궤화공작모식하적평균공작전류분별약위10화300μA ;대10명남성재휴게상태하적심솔진행3 h적련속동태감측,장측시결과여표준동태심전도기록적수거진행비교,결과표명,심솔측량적산술평균오차약위0.3 BPM .
A new low‐power wrist‐worn miniature device used for real‐time wireless heart rate (HR) monitoring was presented . A novel pulse signal detection method based on double channels of pulse sensing of the radial artery and ulnar artery as well as their differential signal was proposed .A miniature device consisting of sensors ,signal condition system was fabricated .The micro‐controller is responsible for power administration ,and calculation of auto gain control algorithm and heart rate algorithms .The average currents are about 10μA and 300μA in standby and active modes ,respectively .Ten male subjects were selected to test the performance of the device through a three‐hour continuous test in their resting condition with the reference of a Holter monitor .The HR of the device was compared with the HR from the ECG signal recorded by the Holter monitor .The experimental results indicate that the arithmetical average error is about 0 .3 BPM .