武警医学
武警醫學
무경의학
MEDICAL JOURNAL OF THE CHINESE PEOPLE'S ARMED POLICE FORCES
2015年
5期
496-498,502
,共4页
荧光内镜%宽视场%近红外荧光探针%肿瘤检测
熒光內鏡%寬視場%近紅外熒光探針%腫瘤檢測
형광내경%관시장%근홍외형광탐침%종류검측
Epi-fluorescence endoscopy%wide-field%near-infrared fluorescent probe%tumor detection
目的:为探索胃癌分子成像技术的新方法,构建一套新型灵活的宽视场荧光内镜系统,并用于裸鼠胃癌移植瘤模型荧光成像研究。方法通过合理的光路设计及光学组件机加工,装配宽视场荧光内镜系统,通过对不同浓度的2-DG-750荧光探针进行成像,分析系统性能;对尾静脉注射100μl 2-DG-750荧光探针的裸鼠胃癌移植瘤模型进行在体荧光成像,对ROI的像素值和成像图进行分析。结果体外成像ROI的像素值与荧光探针浓度存在较好的线性回归关系,在2-DG-750剂量为31.3 pmol时,该系统即可检测到荧光信号,提示该系统灵敏性好;在体成像显示用该系统联合2-DG-750荧光探针观察到了肿瘤组织的较高荧光信号强度。结论宽视场荧光内镜系统具有很好的灵敏性,可以同时实现白光及荧光的双模式成像,具有实现光学分子成像诊断胃癌的潜在应用价值。
目的:為探索胃癌分子成像技術的新方法,構建一套新型靈活的寬視場熒光內鏡繫統,併用于裸鼠胃癌移植瘤模型熒光成像研究。方法通過閤理的光路設計及光學組件機加工,裝配寬視場熒光內鏡繫統,通過對不同濃度的2-DG-750熒光探針進行成像,分析繫統性能;對尾靜脈註射100μl 2-DG-750熒光探針的裸鼠胃癌移植瘤模型進行在體熒光成像,對ROI的像素值和成像圖進行分析。結果體外成像ROI的像素值與熒光探針濃度存在較好的線性迴歸關繫,在2-DG-750劑量為31.3 pmol時,該繫統即可檢測到熒光信號,提示該繫統靈敏性好;在體成像顯示用該繫統聯閤2-DG-750熒光探針觀察到瞭腫瘤組織的較高熒光信號彊度。結論寬視場熒光內鏡繫統具有很好的靈敏性,可以同時實現白光及熒光的雙模式成像,具有實現光學分子成像診斷胃癌的潛在應用價值。
목적:위탐색위암분자성상기술적신방법,구건일투신형령활적관시장형광내경계통,병용우라서위암이식류모형형광성상연구。방법통과합리적광로설계급광학조건궤가공,장배관시장형광내경계통,통과대불동농도적2-DG-750형광탐침진행성상,분석계통성능;대미정맥주사100μl 2-DG-750형광탐침적라서위암이식류모형진행재체형광성상,대ROI적상소치화성상도진행분석。결과체외성상ROI적상소치여형광탐침농도존재교호적선성회귀관계,재2-DG-750제량위31.3 pmol시,해계통즉가검측도형광신호,제시해계통령민성호;재체성상현시용해계통연합2-DG-750형광탐침관찰도료종류조직적교고형광신호강도。결론관시장형광내경계통구유흔호적령민성,가이동시실현백광급형광적쌍모식성상,구유실현광학분자성상진단위암적잠재응용개치。
Objective To setup a new flexible wide-field epi-fluorescence endoscope system, then make a fluorescence ima-ging for gastric cancer xenograft model in nude mice by this system, so as to setup a new device and method for the diagnosis of gastric cancer molecular imaging techniques.Methods The widefield fluorescence endoscope system was assembledusing the rational optical path and optical components machining.Different concentrations of 2-DG-750 fluorescent probe was used to analyze the performance of the imaging system.By intravenous injection of 100μl2-DG-750 fluorescent probe into the xenografted tumor model,in vivo fluorescence imaging wasconducted, and ROI pixel values and imaging map were analyzed.Results The ROI pixel values demonstrated that a sen-sitivity of 31.3 pmoles 2-DG-750 was achieved as well as a good linear response from 1nmole to zero fluorescent probe.Fluorescence imaging indicated that gastrointestinal tumor tissues have an eminent 2-DG-750 uptake yielding a relatively strong fluorescence signal. Conclusions The new flexible wide-field epi-fluorescence endoscope system not only has a good sensitivity, but also can do a dual white and fluorescence imaging, which makes a promise future that optical technique will be applied to the diagnosis of gastric cancer.