实用口腔医学杂志
實用口腔醫學雜誌
실용구강의학잡지
JOURNAL OF PRACTICAL STOMATOLOGY
2015年
3期
313-317
,共5页
党玉琪%张春宝%王伟娜%高勃
黨玉琪%張春寶%王偉娜%高勃
당옥기%장춘보%왕위나%고발
选择性激光熔覆(SLM)%冠%适合性
選擇性激光鎔覆(SLM)%冠%適閤性
선택성격광용복(SLM)%관%괄합성
Selective laser melting(SLM)%Crowns%Fit
目的:评价选择性激光熔覆(SLM)技术与失蜡铸造法制作的钴铬合金单冠与三联冠的适合性。方法:采用2种 SLM系统 EOS M270与 EOS M280及铸造法分别制作钴铬合金单冠与三联冠各5个,通过采用硅橡胶法在体式显微镜下测量边缘及内部垂直间隙。采用 SPSS 17.0软件对实验数据进行单因素方差分析。结果:钴铬合金单冠的边缘间隙和内部间隙(μm)分别为:EOS M270组42.28±11.58、125.75±47.67,EOS M280组46.54±8.4、128.22±54.18,铸造组为48.66±12.08、135.37±46.89。三联冠的边缘间隙(μm)和内部间隙(μm)分别为:EOS M270组40.55±10.04、133.09±39.80,EOS M280组45.36±10、138.94±50.61,铸造组为54.24±12.04、151.87±61.94(P <0.05)。结论:SLM组边缘平均垂直间隙明显小于临床普遍接受的120μm 的标准,精度可以满足临床要求。SLM钴铬合金单冠的边缘适合性优于铸造法,内部适合性无统计学差异。SLM钴铬合金三联冠的边缘适合性及内部适合性均优于铸造法。
目的:評價選擇性激光鎔覆(SLM)技術與失蠟鑄造法製作的鈷鉻閤金單冠與三聯冠的適閤性。方法:採用2種 SLM繫統 EOS M270與 EOS M280及鑄造法分彆製作鈷鉻閤金單冠與三聯冠各5箇,通過採用硅橡膠法在體式顯微鏡下測量邊緣及內部垂直間隙。採用 SPSS 17.0軟件對實驗數據進行單因素方差分析。結果:鈷鉻閤金單冠的邊緣間隙和內部間隙(μm)分彆為:EOS M270組42.28±11.58、125.75±47.67,EOS M280組46.54±8.4、128.22±54.18,鑄造組為48.66±12.08、135.37±46.89。三聯冠的邊緣間隙(μm)和內部間隙(μm)分彆為:EOS M270組40.55±10.04、133.09±39.80,EOS M280組45.36±10、138.94±50.61,鑄造組為54.24±12.04、151.87±61.94(P <0.05)。結論:SLM組邊緣平均垂直間隙明顯小于臨床普遍接受的120μm 的標準,精度可以滿足臨床要求。SLM鈷鉻閤金單冠的邊緣適閤性優于鑄造法,內部適閤性無統計學差異。SLM鈷鉻閤金三聯冠的邊緣適閤性及內部適閤性均優于鑄造法。
목적:평개선택성격광용복(SLM)기술여실사주조법제작적고락합금단관여삼련관적괄합성。방법:채용2충 SLM계통 EOS M270여 EOS M280급주조법분별제작고락합금단관여삼련관각5개,통과채용규상효법재체식현미경하측량변연급내부수직간극。채용 SPSS 17.0연건대실험수거진행단인소방차분석。결과:고락합금단관적변연간극화내부간극(μm)분별위:EOS M270조42.28±11.58、125.75±47.67,EOS M280조46.54±8.4、128.22±54.18,주조조위48.66±12.08、135.37±46.89。삼련관적변연간극(μm)화내부간극(μm)분별위:EOS M270조40.55±10.04、133.09±39.80,EOS M280조45.36±10、138.94±50.61,주조조위54.24±12.04、151.87±61.94(P <0.05)。결론:SLM조변연평균수직간극명현소우림상보편접수적120μm 적표준,정도가이만족림상요구。SLM고락합금단관적변연괄합성우우주조법,내부괄합성무통계학차이。SLM고락합금삼련관적변연괄합성급내부괄합성균우우주조법。
Objective:To compare the adaption of cobalt-chromium prostheses fabricated by selective laser melting(SLM)and conven-tional lost wax technique(LW)method.Methods:Cobalt-chromium crowns and three-unit fixed dental prostheses(FDPs)were fabrica-ted by selective laser melting(SLM)or by conventional lost wax technique(LW)respectively(n =5).Marginal fit and internal fit were examined using a light-body silicone.After setting,each silicon film was cut into 2 parts and the thickness of silicon layer was measured at 1 00 ×magnification using a digital microscope,the data of marginal gap(MG)and internal gap(IG)were statistically analysed by ANOVA and SPSS statistical package version 1 7.0.Results:The MG and IG(μm)of cobalt-chromium alloy crowns in EOS M270 group were 42.28 ±1 1 .58 and 1 25.75 ±47.67,in EOS M280 group 46.54 ±8.4 and 1 28.22 ±54.1 8,in LW group 48.66 ±1 2.08 and 1 35.37 ±46.89.The MG and IG(μm)of three-unit FDPs in EOS M270 group were 40.55 ±1 0.04 and 1 33.09 ±39.80,in EOS M280 group 45.36 ±1 0 and 1 38.94 ±50.61 ,in LW group 54.24 ±1 2.04 and 1 51 .87 ±61 .94(P <0.05),respectively.Conclu-sion:The marginal fit of the SLMgroup is superior to that of the LW group and significantly smaller than 1 20 μm of clinical generally accepted standards.SLMsystem can satisfy the clinical requirements.