非甲烷总烃%甲烷%气相色谱
非甲烷總烴%甲烷%氣相色譜
비갑완총경%갑완%기상색보
NMHC%Methane%GC Method
采用带有双柱单氢焰离子化检测器的气相色谱仪,直接进样差量法测定非甲烷总烃的含量。方法检出限为0.015mg/m3;RSD在0.9%~3.2%之间;加标回收率为95.4%~101.7%。该方法操作简单,定量准确,结果符合实验要求。
採用帶有雙柱單氫燄離子化檢測器的氣相色譜儀,直接進樣差量法測定非甲烷總烴的含量。方法檢齣限為0.015mg/m3;RSD在0.9%~3.2%之間;加標迴收率為95.4%~101.7%。該方法操作簡單,定量準確,結果符閤實驗要求。
채용대유쌍주단경염리자화검측기적기상색보의,직접진양차량법측정비갑완총경적함량。방법검출한위0.015mg/m3;RSD재0.9%~3.2%지간;가표회수솔위95.4%~101.7%。해방법조작간단,정량준학,결과부합실험요구。
Direct injection dispersion method was for determining the content of NMHC. It was determined by GC which has double column and single hydrogen flame ionization detector. Method detection limit was 0.015 mg/m3; RSD was between 0.9%~3.2%. Standard addition recovery was between 95.4%~101.7%. The method was simple operation and could quantify accurately. The results was conform to the experimental requirements.