目的探讨不同 SonoVue 剂量对健康实验兔超声造影(CEUS)效果的影响,并分析其变化规律,探讨最佳 SonoVue 剂量。材料与方法采用 GE LOGIQ-E9超声诊断仪,分别对10例兔肾在8个 SonoVue 剂量水平(0.02~0.16 ml/kg)行 CEUS检查,通过时间-强度曲线测量肾血流灌注参数,并进行定量分析。结果 SonoVue在0.02~0.10 ml/kg 剂量范围内,峰值强度(PI)及曲线下面积(AUC)随剂量增加而增大,SonoVue 剂量与 PI、AUC 呈显著正相关(r=0.962、0.965,P<0.05);进一步增加剂量,AUC 变化不明显,PI 反呈略下降趋势。到达时间(AT)随 SonoVue剂量的增加而呈逐渐缩短,两者呈显著负相关(r=-0.917,P<0.05);SonoVue 剂量与达峰时间(TTP)呈正相关(r=0.49,P<0.05)。结论 SonoVue 剂量会影响CEUS 定量分析结果,确定剂量和效果的关系及认识其规律是进行 CEUS 和定量分析的前提条件。应用 SonoVue 对健康实验兔肾行 CEUS 定量分析时,0.10 ml/kg 是较理想的剂量。
目的探討不同 SonoVue 劑量對健康實驗兔超聲造影(CEUS)效果的影響,併分析其變化規律,探討最佳 SonoVue 劑量。材料與方法採用 GE LOGIQ-E9超聲診斷儀,分彆對10例兔腎在8箇 SonoVue 劑量水平(0.02~0.16 ml/kg)行 CEUS檢查,通過時間-彊度麯線測量腎血流灌註參數,併進行定量分析。結果 SonoVue在0.02~0.10 ml/kg 劑量範圍內,峰值彊度(PI)及麯線下麵積(AUC)隨劑量增加而增大,SonoVue 劑量與 PI、AUC 呈顯著正相關(r=0.962、0.965,P<0.05);進一步增加劑量,AUC 變化不明顯,PI 反呈略下降趨勢。到達時間(AT)隨 SonoVue劑量的增加而呈逐漸縮短,兩者呈顯著負相關(r=-0.917,P<0.05);SonoVue 劑量與達峰時間(TTP)呈正相關(r=0.49,P<0.05)。結論 SonoVue 劑量會影響CEUS 定量分析結果,確定劑量和效果的關繫及認識其規律是進行 CEUS 和定量分析的前提條件。應用 SonoVue 對健康實驗兔腎行 CEUS 定量分析時,0.10 ml/kg 是較理想的劑量。
목적탐토불동 SonoVue 제량대건강실험토초성조영(CEUS)효과적영향,병분석기변화규률,탐토최가 SonoVue 제량。재료여방법채용 GE LOGIQ-E9초성진단의,분별대10례토신재8개 SonoVue 제량수평(0.02~0.16 ml/kg)행 CEUS검사,통과시간-강도곡선측량신혈류관주삼수,병진행정량분석。결과 SonoVue재0.02~0.10 ml/kg 제량범위내,봉치강도(PI)급곡선하면적(AUC)수제량증가이증대,SonoVue 제량여 PI、AUC 정현저정상관(r=0.962、0.965,P<0.05);진일보증가제량,AUC 변화불명현,PI 반정략하강추세。도체시간(AT)수 SonoVue제량적증가이정축점축단,량자정현저부상관(r=-0.917,P<0.05);SonoVue 제량여체봉시간(TTP)정정상관(r=0.49,P<0.05)。결론 SonoVue 제량회영향CEUS 정량분석결과,학정제량화효과적관계급인식기규률시진행 CEUS 화정량분석적전제조건。응용 SonoVue 대건강실험토신행 CEUS 정량분석시,0.10 ml/kg 시교이상적제량。
Purpose To study the effects of different doses of ultrasound contrast agent SonoVue upon contrast-enhanced ultrasound (CEUS) in the kidney of healthy rabbits, and to seek the optimal dose of SonoVue. Materials and Methods CEUS was performed in 10 healthy rabbits with GE LOGIQ-E9 by using 8 different doses of SonoVue (ranging from 0.02 ml/kg to 0.16 ml/kg). The quantitative parameters of the time-intensity curve (TIC) were measured and statistically compared. Results The TICs showed that the peak intensity (PI) and the area under curve (AUC) increased with dose when the doses ranged from 0.02 ml/kg to 0.10 ml/kg (r=0.962 and 0.965, P<0.05); when the dose further increased, AUC had little change but PI decreased reversely. The arrival time (AT) shortened along with the increase of the SonoVue dose (r= - 0.917, P<0.05). The dose had a positive correlation with time to peak (r=0.49, P<0.05). Conclusion The parameters of TIC are influenced intensely by different doses of SonoVue. It is important to realize the relationship between SonoVue doses and its effects upon contrast-enhancement ultrasound in microcirculatory quantification. The dose of 0.10 ml/kg appears to be the optimal dose for CEUS in examining kidney of healthy rabbit.