尹知训%丁红梅%白波%卢伟杰
尹知訓%丁紅梅%白波%盧偉傑
윤지훈%정홍매%백파%로위걸
假体%计算机辅助设计%髋关节
假體%計算機輔助設計%髖關節
가체%계산궤보조설계%관관절
Prosthesis%Computer-aided design%Hip joint
目的:设计出偏心改向髋臼内衬,以解决臼杯位置不良时内衬补救或翻修,及小髋臼关节置换问题。方法在Solidworks2012软件平台上,利用三维实体建模技术,设计一套偏心改向髋臼内衬,并进行模拟手术,预测手术效果。结果成功设计出一套偏心改向髋臼内衬,改向规格有0°、10°、20°、30°;偏心规格有球心偏向非负重侧2 mm的外径/内径的比值分别为42 mm/28 mm、40 mm/28 mm和38 mm/24 mm、36 mm/24 mm四种。模拟手术显示髋关节置换或翻修术后防脱位效果和活动范围。结论偏心改向髋臼内衬设计合理,用于臼杯位置不良时内衬补救或翻修及小髋臼大球头置换,模拟手术提示有防脱位功能。
目的:設計齣偏心改嚮髖臼內襯,以解決臼杯位置不良時內襯補救或翻脩,及小髖臼關節置換問題。方法在Solidworks2012軟件平檯上,利用三維實體建模技術,設計一套偏心改嚮髖臼內襯,併進行模擬手術,預測手術效果。結果成功設計齣一套偏心改嚮髖臼內襯,改嚮規格有0°、10°、20°、30°;偏心規格有毬心偏嚮非負重側2 mm的外徑/內徑的比值分彆為42 mm/28 mm、40 mm/28 mm和38 mm/24 mm、36 mm/24 mm四種。模擬手術顯示髖關節置換或翻脩術後防脫位效果和活動範圍。結論偏心改嚮髖臼內襯設計閤理,用于臼杯位置不良時內襯補救或翻脩及小髖臼大毬頭置換,模擬手術提示有防脫位功能。
목적:설계출편심개향관구내츤,이해결구배위치불량시내츤보구혹번수,급소관구관절치환문제。방법재Solidworks2012연건평태상,이용삼유실체건모기술,설계일투편심개향관구내츤,병진행모의수술,예측수술효과。결과성공설계출일투편심개향관구내츤,개향규격유0°、10°、20°、30°;편심규격유구심편향비부중측2 mm적외경/내경적비치분별위42 mm/28 mm、40 mm/28 mm화38 mm/24 mm、36 mm/24 mm사충。모의수술현시관관절치환혹번수술후방탈위효과화활동범위。결론편심개향관구내츤설계합리,용우구배위치불량시내츤보구혹번수급소관구대구두치환,모의수술제시유방탈위공능。
Objective To design a new type of steerable eccentric acetabular liner for liner remedy or revision when the acetabular cup is in poor position , and for small acetabulum replacement . Methods Using the Solidworks 2012 software, the three dimentional (3D) modeling technology was employed to design a new type of steerable eccentric acetabular liner .The surgical simulations of revision and replacement with the new acetabular liner were performed for predicting the operation effects .Results A new type of steerable eccentric acetabular liner was successfully designed . Four specifications for direction changing included 0°, 10°, 20°, 30°;the four sepcifications of external/internal diameters of eccentric center in the 2 mm-non-weight-bearing side liner were 42/28 mm, 40/28 mm, 38/24 mm, and 36/24 mm. The surgical simulation showed the effects of the hip replacement or revision avoiding dislocation and the range of motions .Conclusion The design of steerable eccentric acetabular liners is reasonable for the liner remedy or revision when the acetabular cup is in poor position and the replacement in patients with small acetabulum; surgical simulation suggests that the new acetabular liners can prevent the artificial joint dislocation .