酸度计%示值%准确度%稳定性
痠度計%示值%準確度%穩定性
산도계%시치%준학도%은정성
acidity meter%indication%accuracy%stablity
对影响酸度计示值准确性及稳定性的因素进行分析。影响酸度计示值准确性及稳定性的因素为玻璃电极被污染、酸度计使用前未校准或校准错误、样品溶液不均匀、酸度计温度设置与样品溶液温度不一致。对影响酸度计示值准确性及稳定性问题提出了解决方法。
對影響痠度計示值準確性及穩定性的因素進行分析。影響痠度計示值準確性及穩定性的因素為玻璃電極被汙染、痠度計使用前未校準或校準錯誤、樣品溶液不均勻、痠度計溫度設置與樣品溶液溫度不一緻。對影響痠度計示值準確性及穩定性問題提齣瞭解決方法。
대영향산도계시치준학성급은정성적인소진행분석。영향산도계시치준학성급은정성적인소위파리전겁피오염、산도계사용전미교준혹교준착오、양품용액불균균、산도계온도설치여양품용액온도불일치。대영향산도계시치준학성급은정성문제제출료해결방법。
The factors affecting the stability and accuracy of acidity meter were analyzed. These factors were as follows:glass electrode contamination,acidity meter calibration or calibration errors before use,uneven sample solution, acidity meter temperature settings were not consistent with the sample solution temperature. These solution methods to solve the problems of accuracy and stability were put forward.