检验医学与临床
檢驗醫學與臨床
검험의학여림상
JOURNAL OF LABORATORY MEDICINE AND CLINICAL SCIENCES
2015年
16期
2421-2422
,共2页
全自动生化分析仪%静脉血清血糖浓度%快速血糖仪%静脉全血血糖浓度
全自動生化分析儀%靜脈血清血糖濃度%快速血糖儀%靜脈全血血糖濃度
전자동생화분석의%정맥혈청혈당농도%쾌속혈당의%정맥전혈혈당농도
automatic biochemical analyzer%blood glucose concentration of venous serum%rapid blood sugar meter%blood glucose concentration of venous w hole blood
目的:探讨全自动生化仪测定静脉促凝血清标本血糖浓度与快速血糖仪测定静脉抗凝全血标本血糖浓度的相关性及差异。方法采集该院200名住院患者的血液标本,同一患者同时采集静脉促凝血清标本与静脉抗凝全血标本,采用全自动生化仪测定静脉促凝血清标本血糖浓度,同时采用快速血糖仪测定静脉抗凝全血标本血糖浓度,比较2组血糖浓度测定结果。结果191名患者的静脉促凝血清标本血糖浓度明显大于其静脉抗凝全血标本血糖浓度,差异范围在0.50~1.50 mmol/L ,差异有统计学意义(P<0.05)。结论全自动生化仪测定静脉促凝血清标本血糖浓度明显高于快速血糖仪测定静脉抗凝全血标本的血糖浓度。
目的:探討全自動生化儀測定靜脈促凝血清標本血糖濃度與快速血糖儀測定靜脈抗凝全血標本血糖濃度的相關性及差異。方法採集該院200名住院患者的血液標本,同一患者同時採集靜脈促凝血清標本與靜脈抗凝全血標本,採用全自動生化儀測定靜脈促凝血清標本血糖濃度,同時採用快速血糖儀測定靜脈抗凝全血標本血糖濃度,比較2組血糖濃度測定結果。結果191名患者的靜脈促凝血清標本血糖濃度明顯大于其靜脈抗凝全血標本血糖濃度,差異範圍在0.50~1.50 mmol/L ,差異有統計學意義(P<0.05)。結論全自動生化儀測定靜脈促凝血清標本血糖濃度明顯高于快速血糖儀測定靜脈抗凝全血標本的血糖濃度。
목적:탐토전자동생화의측정정맥촉응혈청표본혈당농도여쾌속혈당의측정정맥항응전혈표본혈당농도적상관성급차이。방법채집해원200명주원환자적혈액표본,동일환자동시채집정맥촉응혈청표본여정맥항응전혈표본,채용전자동생화의측정정맥촉응혈청표본혈당농도,동시채용쾌속혈당의측정정맥항응전혈표본혈당농도,비교2조혈당농도측정결과。결과191명환자적정맥촉응혈청표본혈당농도명현대우기정맥항응전혈표본혈당농도,차이범위재0.50~1.50 mmol/L ,차이유통계학의의(P<0.05)。결론전자동생화의측정정맥촉응혈청표본혈당농도명현고우쾌속혈당의측정정맥항응전혈표본적혈당농도。
Objective To investigate the correlation and difference between the automatic biochemical analyzer for determining the blood glucose concentration in the venous coagulant serum specimens and the fast blood glucose meter for determining the blood glucose concentration in the venous coagulant whole blood specimens .Methods The blood specimens of 200 inpatients in the Geleshan Branch Hospital were collected ,venous coagulant serum specimens and the venous coagulant whole blood specimens in the same patient were simultaneously collected .The automatic bi‐ochemical analyzer was adopted to detect the blood glucose concentration of venous coagulant serum specimens ,and at the same time the rapid blood glucose meter was adopted to detect the blood glucose concentration of venous coagu‐lant whole blood specimens .The blood glucose detection results were compared between the two detection methods . Results In 191 patients ,the blood glucose concentration of venous coagulant serum specimens was significantly grea‐ter than that of the venous coagulant whole blood specimens ,the difference range was 0 .50 -1 .50 mmol/L ,the difference was statistically significant (P<0 .05) .Conclusion The blood glucose concentration of the venous coagu‐lant serum specimens detected by the full automatic biochemical analyzer is significantly higher than that of the raven‐ous coagulant whole blood specimens detected by the fast blood glucose meter .