实用医院临床杂志
實用醫院臨床雜誌
실용의원림상잡지
PRACTICAL JOURNAL OF CLINICAL MEDICINE
2015年
5期
144-146,147
,共4页
杨艳%袁蓉%唐旭%金丹%李蓬秋
楊豔%袁蓉%唐旭%金丹%李蓬鞦
양염%원용%당욱%금단%리봉추
糖尿病%血糖仪%生化分析仪%质量
糖尿病%血糖儀%生化分析儀%質量
당뇨병%혈당의%생화분석의%질량
Diabetes%Glucose meters%Biochemical analyzer%Quality
目的:比较便携式(point of care testing,POCT)血糖仪与生化分析仪的血糖检测结果,探讨 POCT 血糖仪质量控制与管理。方法①应用强生 POCT 血糖仪(强生稳步血糖仪和强生稳豪倍优血糖仪)及全自动生化分析仪检测190例空腹静脉抗凝(EDTA-K2抗凝)全血标本血糖浓度,并进行比较;②检测该仪器的精密度、准确度并判断临床可接受性,分析血糖仪结果与参比检测系统生化分析仪间的相关性。结果两款 POCT 血糖仪的批内变异系数(CV)及批间 CV 均小于5%,差异无统计学意义(P >0.05);190份样本均应用经血糖仪与生化分析仪检测,血糖仪98.9%的检测结果误差在允许误差范围内,临床判断均可接受;POCT 血糖仪检测结果与参比检测系统生化分析仪检测结果呈正相关(r 分别为0.9953和0.9870,均 P <0.05)。结论两种强生 POCT 血糖仪均具有良好的精密度、准确性及线性范围。
目的:比較便攜式(point of care testing,POCT)血糖儀與生化分析儀的血糖檢測結果,探討 POCT 血糖儀質量控製與管理。方法①應用彊生 POCT 血糖儀(彊生穩步血糖儀和彊生穩豪倍優血糖儀)及全自動生化分析儀檢測190例空腹靜脈抗凝(EDTA-K2抗凝)全血標本血糖濃度,併進行比較;②檢測該儀器的精密度、準確度併判斷臨床可接受性,分析血糖儀結果與參比檢測繫統生化分析儀間的相關性。結果兩款 POCT 血糖儀的批內變異繫數(CV)及批間 CV 均小于5%,差異無統計學意義(P >0.05);190份樣本均應用經血糖儀與生化分析儀檢測,血糖儀98.9%的檢測結果誤差在允許誤差範圍內,臨床判斷均可接受;POCT 血糖儀檢測結果與參比檢測繫統生化分析儀檢測結果呈正相關(r 分彆為0.9953和0.9870,均 P <0.05)。結論兩種彊生 POCT 血糖儀均具有良好的精密度、準確性及線性範圍。
목적:비교편휴식(point of care testing,POCT)혈당의여생화분석의적혈당검측결과,탐토 POCT 혈당의질량공제여관리。방법①응용강생 POCT 혈당의(강생은보혈당의화강생은호배우혈당의)급전자동생화분석의검측190례공복정맥항응(EDTA-K2항응)전혈표본혈당농도,병진행비교;②검측해의기적정밀도、준학도병판단림상가접수성,분석혈당의결과여삼비검측계통생화분석의간적상관성。결과량관 POCT 혈당의적비내변이계수(CV)급비간 CV 균소우5%,차이무통계학의의(P >0.05);190빈양본균응용경혈당의여생화분석의검측,혈당의98.9%적검측결과오차재윤허오차범위내,림상판단균가접수;POCT 혈당의검측결과여삼비검측계통생화분석의검측결과정정상관(r 분별위0.9953화0.9870,균 P <0.05)。결론량충강생 POCT 혈당의균구유량호적정밀도、준학성급선성범위。
Objective To compare the testing results between point of care testing (POCT)glucose meters and biochemical an -alyzer,and investigate the quality control and management of POCT glucose meters .Methods Blood glucose concentrations of 190 ve-nous whole blood samples were simultaneously detected by using a Johnson &Johnson portable blood glucose meters (One Touch Surest-ep and One Touch UltraVue)and an automatic biochemical analyzer .The testing results were compared between the two methods .In ad-dition,precision,accuracy and clinical acceptability of the POCT glucose meters were assessed ,and correlation of the results between the two methods was analyzed.Results Coefficients of variations (CVs)of within-and between-batches of two types of POCT glucose meters were all less than 5%.98.9% of the testing results of POCT glucose meters were within the allowable error limits and accepted clinically.POCT glucose meters had a positive correlation with biochemical analyzer in the test results of blood glucose (r =0.9953, 0.9870,respectively,P <0.05).Conclusion The 2 types of POCT glucose meters all have good precision ,accuracy and linearrange. Scheduled comparison with biochemical analyzer and establishing the procedure of quality management play an important role in the quality control of POCT glucose meters.