天津医药
天津醫藥
천진의약
TIANJIN MEDICAL JOURNAL
2015年
9期
1066-1068
,共3页
王林妮%张磊%秦鲁宁%张红
王林妮%張磊%秦魯寧%張紅
왕림니%장뢰%진로저%장홍
角膜%散光%屈光计检查%矢量%自动验光仪%Lenstar%KR-1
角膜%散光%屈光計檢查%矢量%自動驗光儀%Lenstar%KR-1
각막%산광%굴광계검사%시량%자동험광의%Lenstar%KR-1
cornea%astigmatism%refractometry%vector%auto-refractor%Lenstar%KR-1
目的:比较年龄相关性白内障患者术前Lenstar LS900和KR-1全自动验光仪测量角膜曲率及角膜散光各参数的差异。方法收集我院年龄相关性白内障患者76例(76眼),术前分别通过Lenstar LS900和KR-1全自动验光仪测量角膜相关参数,包括平坦角膜曲率(K1)、陡峭角膜曲率(K2)、K1和K2的平均值(Km)、散光值以及散光轴位,其中将角膜散光通过傅立叶矢量转换公式转换成矢量表示法(J0,J45)进行比较。结果 Lenstar LS900测量K1、K2、Km、J0和J45分别为(43.960±1.440)D、(44.901±1.319)D、(44.430±1.336)D、0.043±0.402、0.017±0.425;KR-1全自动验光仪测量K1、K2、Km、J0和J45分别为(44.007±1.400)D、(44.859±1.338)D、(44.433±1.330)D、-0.058±0.322、0.031±0.419;差异均无统计学意义(均P>0.05)。Bland-Altman分析说明两种仪器的检测结果具有95%的一致性界限。结论 Lenstar LS900及KR-1全自动验光仪均可以应用于年龄相关性白内障患者术前测量角膜散光的各项参数。
目的:比較年齡相關性白內障患者術前Lenstar LS900和KR-1全自動驗光儀測量角膜麯率及角膜散光各參數的差異。方法收集我院年齡相關性白內障患者76例(76眼),術前分彆通過Lenstar LS900和KR-1全自動驗光儀測量角膜相關參數,包括平坦角膜麯率(K1)、陡峭角膜麯率(K2)、K1和K2的平均值(Km)、散光值以及散光軸位,其中將角膜散光通過傅立葉矢量轉換公式轉換成矢量錶示法(J0,J45)進行比較。結果 Lenstar LS900測量K1、K2、Km、J0和J45分彆為(43.960±1.440)D、(44.901±1.319)D、(44.430±1.336)D、0.043±0.402、0.017±0.425;KR-1全自動驗光儀測量K1、K2、Km、J0和J45分彆為(44.007±1.400)D、(44.859±1.338)D、(44.433±1.330)D、-0.058±0.322、0.031±0.419;差異均無統計學意義(均P>0.05)。Bland-Altman分析說明兩種儀器的檢測結果具有95%的一緻性界限。結論 Lenstar LS900及KR-1全自動驗光儀均可以應用于年齡相關性白內障患者術前測量角膜散光的各項參數。
목적:비교년령상관성백내장환자술전Lenstar LS900화KR-1전자동험광의측량각막곡솔급각막산광각삼수적차이。방법수집아원년령상관성백내장환자76례(76안),술전분별통과Lenstar LS900화KR-1전자동험광의측량각막상관삼수,포괄평탄각막곡솔(K1)、두초각막곡솔(K2)、K1화K2적평균치(Km)、산광치이급산광축위,기중장각막산광통과부립협시량전환공식전환성시량표시법(J0,J45)진행비교。결과 Lenstar LS900측량K1、K2、Km、J0화J45분별위(43.960±1.440)D、(44.901±1.319)D、(44.430±1.336)D、0.043±0.402、0.017±0.425;KR-1전자동험광의측량K1、K2、Km、J0화J45분별위(44.007±1.400)D、(44.859±1.338)D、(44.433±1.330)D、-0.058±0.322、0.031±0.419;차이균무통계학의의(균P>0.05)。Bland-Altman분석설명량충의기적검측결과구유95%적일치성계한。결론 Lenstar LS900급KR-1전자동험광의균가이응용우년령상관성백내장환자술전측량각막산광적각항삼수。
Objective To compare the differences of corneal keratometry and corneal astigmatism measured by Len?star LS900 and KR-1 auto-refractor of age-related cataract. Methods Seventy-six patients with cataract (76 eyes) were in?cluded in this study. Flat keratometry (K1), steep keratometry (K2), mean keratometry (Km) and astigmatism were measured before operation by Lenstar LS900 and KR-1 auto-refractor. The parameters of astigmatism were transformed into J0 and J45 by Fourier vector transform, and which was compared. Results Values of K1, K2, Km, J0 and J45 were (43.960±1.440) D, (44.901±1.319)D, (44.430±1.336)D, 0.043±0.402 and 0.017±0.425 measured by Lenstar LS900, respectively, and which measured by KR-1 auto-refractor were (44.007 ± 1.400)D, (44.859 ± 1.338)D, (44.433 ± 1.330)D,-0.058 ± 0.322 and 0.031 ± 0.419, respectively. There was no statistical difference between these values measured by two instruments ( P>0.05). The Bland-Altman plots showed that two devices had coincident results for corneal parameters. Conclusion Lenstar LS900 and KR-1 auto-refractor can be applied in the measurement of corneal astigmatism of age-related cataract before surgery.