Journal of Beijing University of Physical Education
2013年
4期
77~81
,共null页
能量消耗 走跑运动 传感器 能量消耗 Livepod LP2 Armband Actigraph GT3X
能量消耗 走跑運動 傳感器 能量消耗 Livepod LP2 Armband Actigraph GT3X
능량소모 주포운동 전감기 능량소모 Livepod LP2 Armband Actigraph GT3X
energy expenditure ; walking and running; accelerometers ; Livepod LP2 ; Armband ; Actigraph GT3X
目的:验证Actigraph GT3X、Livepod LP2、Armband 3种加速度传感器测量人体走、跑运动能量消耗的信效度水平。方法:12名健康男生与14名健康女生佩戴Actigraph GT3X、Livepod LP2、Armband 3种仪器以及与气体代谢分析仪MAXⅡ相连的呼吸面罩,在跑台上完成速度水平分别为dKPH、6KPH和8KPH,每种速度水平下时间为10min的实验。通过计算组内相关系数(ICC值)比较3种仪器测量一致性,以确定其信度水平;采用单因素方差分析比较同一仪器不同速度之间测量值差异度;采用配对样本T检验分析3种仪器测量值与校标MAXII测量值一致性程度;采用相关分析衡量3种仪器测量值与MAXⅡ测量值相关性;综合以上3种方法确定3种仪器的测量效度水平。结果:3种仪器ICC值均大于0.93,所有仪器测量值在同一速度水平上具有良好一致性;不同速度仪器测量值差异具有统计学意义;Livepod LP2与MAXⅡ测量值相关系数介于0.71(P〈0.01)和0.8(P〈0.01)(均大于各速度水平其他两仪器与MAXⅡ测量值相关系数)其二者测量值差异没有统计学意义;所有速度水平上Armband高估能量消耗,除8KPH速度水平外,Actigraph GT3X显著低估运动能量消耗。结论:Actigraph GT3X、Livepod LP2、Armband具有较高的信度水平,在测量人体走跑运动能量消耗时,Livepod LP2具有最高的效度水平。
目的:驗證Actigraph GT3X、Livepod LP2、Armband 3種加速度傳感器測量人體走、跑運動能量消耗的信效度水平。方法:12名健康男生與14名健康女生珮戴Actigraph GT3X、Livepod LP2、Armband 3種儀器以及與氣體代謝分析儀MAXⅡ相連的呼吸麵罩,在跑檯上完成速度水平分彆為dKPH、6KPH和8KPH,每種速度水平下時間為10min的實驗。通過計算組內相關繫數(ICC值)比較3種儀器測量一緻性,以確定其信度水平;採用單因素方差分析比較同一儀器不同速度之間測量值差異度;採用配對樣本T檢驗分析3種儀器測量值與校標MAXII測量值一緻性程度;採用相關分析衡量3種儀器測量值與MAXⅡ測量值相關性;綜閤以上3種方法確定3種儀器的測量效度水平。結果:3種儀器ICC值均大于0.93,所有儀器測量值在同一速度水平上具有良好一緻性;不同速度儀器測量值差異具有統計學意義;Livepod LP2與MAXⅡ測量值相關繫數介于0.71(P〈0.01)和0.8(P〈0.01)(均大于各速度水平其他兩儀器與MAXⅡ測量值相關繫數)其二者測量值差異沒有統計學意義;所有速度水平上Armband高估能量消耗,除8KPH速度水平外,Actigraph GT3X顯著低估運動能量消耗。結論:Actigraph GT3X、Livepod LP2、Armband具有較高的信度水平,在測量人體走跑運動能量消耗時,Livepod LP2具有最高的效度水平。
목적:험증Actigraph GT3X、Livepod LP2、Armband 3충가속도전감기측량인체주、포운동능량소모적신효도수평。방법:12명건강남생여14명건강녀생패대Actigraph GT3X、Livepod LP2、Armband 3충의기이급여기체대사분석의MAXⅡ상련적호흡면조,재포태상완성속도수평분별위dKPH、6KPH화8KPH,매충속도수평하시간위10min적실험。통과계산조내상관계수(ICC치)비교3충의기측량일치성,이학정기신도수평;채용단인소방차분석비교동일의기불동속도지간측량치차이도;채용배대양본T검험분석3충의기측량치여교표MAXII측량치일치성정도;채용상관분석형량3충의기측량치여MAXⅡ측량치상관성;종합이상3충방법학정3충의기적측량효도수평。결과:3충의기ICC치균대우0.93,소유의기측량치재동일속도수평상구유량호일치성;불동속도의기측량치차이구유통계학의의;Livepod LP2여MAXⅡ측량치상관계수개우0.71(P〈0.01)화0.8(P〈0.01)(균대우각속도수평기타량의기여MAXⅡ측량치상관계수)기이자측량치차이몰유통계학의의;소유속도수평상Armband고고능량소모,제8KPH속도수평외,Actigraph GT3X현저저고운동능량소모。결론:Actigraph GT3X、Livepod LP2、Armband구유교고적신도수평,재측량인체주포운동능량소모시,Livepod LP2구유최고적효도수평。
Objective: To evaluate the reliability and validity of the Actigraph GT3X, the Livepod LP2, and the Armband in physical activity monitor. Methods : A total of 12 healthy men and 14 healthy women performed 30 mi- nutes of treadmill walking at 4 km/h and 6 km/h and treadmill running at 8 km/h. The Livepod LP2, Armband, and Actigraph GT3X were placed as recommended and indirect calorimetry was continuously recorded during all exercises. The ICC of three aceelerometers was calculated to measure their reliability, and data were analyzed with one way ANVOA, pairwise t-test, and correlation analysis to measure their validity. Results : The ICC of three ac- celerometers was all higher than 0.93 indicating that every accelerometer was on high reliability level. There was significant difference in mean EE at the speed of 4 KPH, 6 KPH, and 8 KPH for every accelerometer, respective- ly. The correlations between measures of IC and Livepod LP2 were form 0.71 ( P 〈 0.01 ) and 0.8 ( P 〈 0.01 ) which were higher than others. At all speed level, the Armband overestimated EE, on the contrary, the Aetigraph underestimated EE at all speed level except 8 KPH. Conclusion: The Livepod LP2, Armband, and Actigraph GT3X have high measurement reliability level. Livepod LP2 can estimate total EE at walking and running best.