Isomed 2000测试系统 等速肌力 等速向心 组内相关系数
Isomed 2000測試繫統 等速肌力 等速嚮心 組內相關繫數
Isomed 2000측시계통 등속기력 등속향심 조내상관계수
Isomed2000 dynemometer; isokinetic muscle strength; isokinetic concentric; intraclass correlation coefficient
分析Isomed 2000测试系统等速向心肌力测试方法的可信度。应用标准重物对比的方法对Isomed 2000等速肌力测试系统进行检验,对19名普通健康年青人(15男,4女,平均年龄=24.1±5.6岁)进行了腰背屈/伸60°/s、髋关节屈/伸60°/s、180°/s和240°/s、膝关节屈/伸60°/s和240°/s、踝关节跖屈/背屈60°/s、180°/s和240°/s、肩关节60°/s和180°/s、肘关节屈/伸60°/s、180°/s和240°/s、腕关节屈/伸60°/s和240°/s的单侧等速向心测试和重复性测试,2次测试间隔4-7天。测试结果:1)标准重物进行的3次测试值与计算值的相关系数r〉0.999;2)腰背屈/伸60°/s、髋关节屈/伸60°/s和180°/s、膝关节屈/伸60°/s和240°/s、踝关节跖屈/背屈60°/s和180°/s、肩关节60°/s和180°/s、肘关节屈/伸60°/s和180°/s、腕关节屈/伸60°/s和240°/s测试时的力矩、做功和功率参数的组内相关系数(ICC)介于0.71-0.98之间;3)肘关节240°/s测试时力矩的ICC为0.67-0.82、髋关节240°/s测试时屈肌群各参数的ICC为0.64-0.83、踝关节240°/s测试时背屈肌群各参数值的ICC为0.36-0.82。结论:该仪器误差小,准确度、精度良好;肘关节、髋关节和踝关节240°/s测试重复性相对较差,其余各关节肌群的不同速度测试重复性中等或者良好。
分析Isomed 2000測試繫統等速嚮心肌力測試方法的可信度。應用標準重物對比的方法對Isomed 2000等速肌力測試繫統進行檢驗,對19名普通健康年青人(15男,4女,平均年齡=24.1±5.6歲)進行瞭腰揹屈/伸60°/s、髖關節屈/伸60°/s、180°/s和240°/s、膝關節屈/伸60°/s和240°/s、踝關節蹠屈/揹屈60°/s、180°/s和240°/s、肩關節60°/s和180°/s、肘關節屈/伸60°/s、180°/s和240°/s、腕關節屈/伸60°/s和240°/s的單側等速嚮心測試和重複性測試,2次測試間隔4-7天。測試結果:1)標準重物進行的3次測試值與計算值的相關繫數r〉0.999;2)腰揹屈/伸60°/s、髖關節屈/伸60°/s和180°/s、膝關節屈/伸60°/s和240°/s、踝關節蹠屈/揹屈60°/s和180°/s、肩關節60°/s和180°/s、肘關節屈/伸60°/s和180°/s、腕關節屈/伸60°/s和240°/s測試時的力矩、做功和功率參數的組內相關繫數(ICC)介于0.71-0.98之間;3)肘關節240°/s測試時力矩的ICC為0.67-0.82、髖關節240°/s測試時屈肌群各參數的ICC為0.64-0.83、踝關節240°/s測試時揹屈肌群各參數值的ICC為0.36-0.82。結論:該儀器誤差小,準確度、精度良好;肘關節、髖關節和踝關節240°/s測試重複性相對較差,其餘各關節肌群的不同速度測試重複性中等或者良好。
분석Isomed 2000측시계통등속향심기력측시방법적가신도。응용표준중물대비적방법대Isomed 2000등속기력측시계통진행검험,대19명보통건강년청인(15남,4녀,평균년령=24.1±5.6세)진행료요배굴/신60°/s、관관절굴/신60°/s、180°/s화240°/s、슬관절굴/신60°/s화240°/s、과관절척굴/배굴60°/s、180°/s화240°/s、견관절60°/s화180°/s、주관절굴/신60°/s、180°/s화240°/s、완관절굴/신60°/s화240°/s적단측등속향심측시화중복성측시,2차측시간격4-7천。측시결과:1)표준중물진행적3차측시치여계산치적상관계수r〉0.999;2)요배굴/신60°/s、관관절굴/신60°/s화180°/s、슬관절굴/신60°/s화240°/s、과관절척굴/배굴60°/s화180°/s、견관절60°/s화180°/s、주관절굴/신60°/s화180°/s、완관절굴/신60°/s화240°/s측시시적력구、주공화공솔삼수적조내상관계수(ICC)개우0.71-0.98지간;3)주관절240°/s측시시력구적ICC위0.67-0.82、관관절240°/s측시시굴기군각삼수적ICC위0.64-0.83、과관절240°/s측시시배굴기군각삼수치적ICC위0.36-0.82。결론:해의기오차소,준학도、정도량호;주관절、관관절화과관절240°/s측시중복성상대교차,기여각관절기군적불동속도측시중복성중등혹자량호。
The purpose of this study was to evaluate the reliability of Isomed 2000 dynamometer.We compared the dynamometer and different weight of barbells on torque.Nineteen normal subjects(15male,4female)ages 24.1±5.6with no history of related injury were tested unilaterally for concentric extension/flexion,back at 60°/s,hip at 60°/s,180°/s and 240°/s,knee at 60°/s and 240°/s,ankle at 60°/s,180°/s and 240°/s,shoulder at 60°/s and 180°/s,elbow at 60°/s,180°/s and 240°/s,wrist at 60°/s and 240°/s,utilizing standard Isomed2000 protocol.Each subject was tested two times with an interval of 4-7days.Statistical analysis of data showed that 1)The correlation coefficient(r)between the torque measured and calculated was〉0.999.2)the isokinetic concentric extension/flexion mode for back at 60°/s,hip at 60°/s and 180°/s,knee at 60°/s and 240°/s,ankle at 60°/s and 180°/s,shoulder at60°/s and 180°/s,elbow at 60°/s and 180°/s,wrist at 60°/s and 240°/s of the Isomed 2000 dynamometer was with test-retest reliability ranging from 0.71 to 0.98.3)The intraclass correlation coefficient(ICC)of elbow extension/flexion peak torque and average peak torque at240°/s to be from 0.67 to 0.82,ICC of hip flexion every parameters at 240°/s to be from0.64 to 0.83,ICC of ankle extension(dorsiflexion)every parameters at 240°/s to be from0.36 to 0.82.Conclusions:1)The measurement error of the dynamometer is small and the accuracy is fine.2)The isokinetic concentric extension/flexion mode for hip,elbow and ankle at240°/s of the dynamometer was less or not reliable for test-retest measures of peak torque,work and power;3)Other measures implied good to excellent reliable.