目的:评估 Beckman Coulter 实验室自动化系统生化免疫样品间携带污染情况,探讨不选择分杯模块时生化免疫共用一管血的可行性。方法将高浓度及不同样品量(2000μl,500μl)低浓度的 HCG 样品按照一定顺序在 Beckman AU5421全自动生化分析仪上检测生化项目,比较检测生化项目前后低浓度样品 HCG 含量的变化评估生化免疫样品间携带污染情况,并在增加 Beckman AU5421全自动生化分析仪特殊清洗程序后重复试验。结果在不同样品量(2000μl,500μl)条件下,Beckman AU5421全自动生化分析仪单模块及整机携带污染分别为5.44,15.47,23.51,45.96 ppm,不同样品量及单模块与整机间携带污染差异均有统计学意义(t =14.553,P <0.001;t =5.527,P =0.005;t =3.985,P =0.016;t=20.457,P <0.001);增加 Beckman AU5421特殊清洗程序后仅检出500μl 样品量时整机携带污染为0.22 ppm。结论不选择分杯模块时,Beckman Coulter 实验室自动化系统可使用特殊清洗程序解决生化免疫间携带污染。
目的:評估 Beckman Coulter 實驗室自動化繫統生化免疫樣品間攜帶汙染情況,探討不選擇分杯模塊時生化免疫共用一管血的可行性。方法將高濃度及不同樣品量(2000μl,500μl)低濃度的 HCG 樣品按照一定順序在 Beckman AU5421全自動生化分析儀上檢測生化項目,比較檢測生化項目前後低濃度樣品 HCG 含量的變化評估生化免疫樣品間攜帶汙染情況,併在增加 Beckman AU5421全自動生化分析儀特殊清洗程序後重複試驗。結果在不同樣品量(2000μl,500μl)條件下,Beckman AU5421全自動生化分析儀單模塊及整機攜帶汙染分彆為5.44,15.47,23.51,45.96 ppm,不同樣品量及單模塊與整機間攜帶汙染差異均有統計學意義(t =14.553,P <0.001;t =5.527,P =0.005;t =3.985,P =0.016;t=20.457,P <0.001);增加 Beckman AU5421特殊清洗程序後僅檢齣500μl 樣品量時整機攜帶汙染為0.22 ppm。結論不選擇分杯模塊時,Beckman Coulter 實驗室自動化繫統可使用特殊清洗程序解決生化免疫間攜帶汙染。
목적:평고 Beckman Coulter 실험실자동화계통생화면역양품간휴대오염정황,탐토불선택분배모괴시생화면역공용일관혈적가행성。방법장고농도급불동양품량(2000μl,500μl)저농도적 HCG 양품안조일정순서재 Beckman AU5421전자동생화분석의상검측생화항목,비교검측생화항목전후저농도양품 HCG 함량적변화평고생화면역양품간휴대오염정황,병재증가 Beckman AU5421전자동생화분석의특수청세정서후중복시험。결과재불동양품량(2000μl,500μl)조건하,Beckman AU5421전자동생화분석의단모괴급정궤휴대오염분별위5.44,15.47,23.51,45.96 ppm,불동양품량급단모괴여정궤간휴대오염차이균유통계학의의(t =14.553,P <0.001;t =5.527,P =0.005;t =3.985,P =0.016;t=20.457,P <0.001);증가 Beckman AU5421특수청세정서후부검출500μl 양품량시정궤휴대오염위0.22 ppm。결론불선택분배모괴시,Beckman Coulter 실험실자동화계통가사용특수청세정서해결생화면역간휴대오염。
Objective To evaluate the Carryover between the chemistry and immunoassay in Beckman Coulter Laboratory Au-tomation System and decide to whether sharing samples for testing between chemistry and immunoassay systems or not. Methods According to a certain order,high concentration samples and low concentration samples of HCG with different sample volume (500 μl,2 000 μl)were tested on Beckman AU5421 automatic biochemical analyzer.The HCG of low concen-tration samples were then tested to evaluate the carryover between the chemistry and immunoassay and explored the correc-tive procedure to deal with the carryover by increasing special cleaning process of beckman AU5421 automatic biochemistry analyzer.Results Under different sample volume,the carryover in a single module and as a whole of the beckman AU5421 automatic biochemistry analyzer were 5.44,15.47,23.51 and 45.96 ppm respectively (t=14.553,P <0.001;t=5.527,P =0.005;t=3.985,P =0.016;t=20.457,P <0.001).By increasing special cleaning process the carryover of 0.22 ppm was detected in 500 μl sample volume of the beckman AU5421 automatic biochemistry analyzer as a whole.Conclusion The car-ryover between the chemistry and immunoassay in Beckman Coulter Laboratory Automation System could been sovled by in-creasing special cleaning process of beckman AU5421 automatic biochemistry analyzer.