目的:研究环境条件对注射泵质量性能检测的影响,为进行注射泵现场质量检测提供依据。方法采用现场运行试验方法,分别在高温、湿热、低温和存在振动环境条件下,对注射泵流速准确性和阻塞报警阈值进行测试,将测量结果与标准环境条件下的测量结果对比,对照分析气候条件和振动对注射泵性能检测的影响规律。结果在高温和湿热环境下,流量测量结果偏高,输出相对误差明显增加,超出最大允许误差;低温环境会使流量测量结果偏低,输出相对误差明显增加,超出最大允许误差;振动对测量结果无明显影响;环境条件对阻塞报警压力阈值测量结果无影响。结论必须加强注射泵现场检测技术的研究。
目的:研究環境條件對註射泵質量性能檢測的影響,為進行註射泵現場質量檢測提供依據。方法採用現場運行試驗方法,分彆在高溫、濕熱、低溫和存在振動環境條件下,對註射泵流速準確性和阻塞報警閾值進行測試,將測量結果與標準環境條件下的測量結果對比,對照分析氣候條件和振動對註射泵性能檢測的影響規律。結果在高溫和濕熱環境下,流量測量結果偏高,輸齣相對誤差明顯增加,超齣最大允許誤差;低溫環境會使流量測量結果偏低,輸齣相對誤差明顯增加,超齣最大允許誤差;振動對測量結果無明顯影響;環境條件對阻塞報警壓力閾值測量結果無影響。結論必鬚加彊註射泵現場檢測技術的研究。
목적:연구배경조건대주사빙질량성능검측적영향,위진행주사빙현장질량검측제공의거。방법채용현장운행시험방법,분별재고온、습열、저온화존재진동배경조건하,대주사빙류속준학성화조새보경역치진행측시,장측량결과여표준배경조건하적측량결과대비,대조분석기후조건화진동대주사빙성능검측적영향규률。결과재고온화습열배경하,류량측량결과편고,수출상대오차명현증가,초출최대윤허오차;저온배경회사류량측량결과편저,수출상대오차명현증가,초출최대윤허오차;진동대측량결과무명현영향;배경조건대조새보경압력역치측량결과무영향。결론필수가강주사빙현장검측기술적연구。
Objective To explore the inlfuence of environmental conditions on injection pump performance testing so as to provide foundations for field quality testing of injection pumps. Methods By using the field testing method, the flow accuracy and blocking alarm threshold of the injection pump were detected respectively under the high temperature, damp heat, low temperature and vibration environment conditions. The testing results under field environment were compared with those under laboratory environment so as to study the influence of climatic conditions and vibration on the injection pump performance testing. Results Under the high temperature and humid environment, lfow measurement was higher and the relative error of the flow was beyond the maximum permissible error. Under the low temperature environment, the lfow measurement was lower and the relative error of the lfow was beyond the maximum permissible error. Vibration had no obvious inlfuence on the testing results. The blocking alarm threshold was not affected by environment. Conclusion Study on ifeld testing technique of injection pumps must be strengthened.