2015年
10期
124-126
,共3页
血细胞分析仪%白细胞计数%红细胞计数%血红蛋白%红细胞比积%血小板计数
血細胞分析儀%白細胞計數%紅細胞計數%血紅蛋白%紅細胞比積%血小闆計數
혈세포분석의%백세포계수%홍세포계수%혈홍단백%홍세포비적%혈소판계수
blood cell analyzer%white blood cell count%red blood cell count%hemoglobin%red blood cells than product%platelet count
目的:通过比较MEK7222及Sysmex XT-1800i全自动五分类血细胞分析仪的检测结果,观察两台仪器的相关性和一致性。方法统计当月质控结果数据,分析在控情况,保证两台仪器性能处于最佳状态;选取61例患者,每例患者取两份血液样本,分别通过MEK7222和XT-1800i分析仪进行检测,观察两台仪器对白细胞计数(WBC)、红细胞计数(RBC)、血红蛋白(HBG)、红细胞比积(HCT)、血小板计数(PLT)的检测结果在统计学意义上的相关性。结果两台仪器所测5项指标的结果在统计学上无明显差异(P>0.05),WBC、RBC、HBG、HCT及PLT的相关系数分别为0.99、0.98、0.99、0.99、0.97,表明两组结果相关性良好。结论 MEK7222和XT-1800i分析仪检测结果有较高的一致性和相关性,适合在同一实验室应用。
目的:通過比較MEK7222及Sysmex XT-1800i全自動五分類血細胞分析儀的檢測結果,觀察兩檯儀器的相關性和一緻性。方法統計噹月質控結果數據,分析在控情況,保證兩檯儀器性能處于最佳狀態;選取61例患者,每例患者取兩份血液樣本,分彆通過MEK7222和XT-1800i分析儀進行檢測,觀察兩檯儀器對白細胞計數(WBC)、紅細胞計數(RBC)、血紅蛋白(HBG)、紅細胞比積(HCT)、血小闆計數(PLT)的檢測結果在統計學意義上的相關性。結果兩檯儀器所測5項指標的結果在統計學上無明顯差異(P>0.05),WBC、RBC、HBG、HCT及PLT的相關繫數分彆為0.99、0.98、0.99、0.99、0.97,錶明兩組結果相關性良好。結論 MEK7222和XT-1800i分析儀檢測結果有較高的一緻性和相關性,適閤在同一實驗室應用。
목적:통과비교MEK7222급Sysmex XT-1800i전자동오분류혈세포분석의적검측결과,관찰량태의기적상관성화일치성。방법통계당월질공결과수거,분석재공정황,보증량태의기성능처우최가상태;선취61례환자,매례환자취량빈혈액양본,분별통과MEK7222화XT-1800i분석의진행검측,관찰량태의기대백세포계수(WBC)、홍세포계수(RBC)、혈홍단백(HBG)、홍세포비적(HCT)、혈소판계수(PLT)적검측결과재통계학의의상적상관성。결과량태의기소측5항지표적결과재통계학상무명현차이(P>0.05),WBC、RBC、HBG、HCT급PLT적상관계수분별위0.99、0.98、0.99、0.99、0.97,표명량조결과상관성량호。결론 MEK7222화XT-1800i분석의검측결과유교고적일치성화상관성,괄합재동일실험실응용。
Objective Through comparisons between test results of two fully-automatic ifve classiifcation blood cell analyzers to observe their correlation and consistency.Methods Statistical data of the quality control results during the month were analyzed to ensure the performance of two instruments was in the best state. And then 61 patients were randomly selected. Two blood samples were taken from each patient and tested by MEK7222 and XT–1800i Analyzer to observe the statistically signiifcant correlation of ifve indicators, including white blood cell count (WBC), red blood cell count (RBC), hemoglobin (HBG), red blood cells than product (HCT) and platelet count (PLT).Results There were no signiifcant differences in the 5 indicators between two instruments (P>0.05). The correlation coefifcients of WBC, RBC, HBG, WBC, HCT and PLT were 0.99, 0.98, 0.99, 0.99 and 0.97, respectively, which showed the existence of correlation between the two groups.Conclusion According to the test results, MEK7222 and XT–1800i Analyzer showed high consistency and correlation, which were suitable for application in the same laboratory.