硅藻土助滤剂%电感耦合等离子体质谱%重金属%内标%动态反应池
硅藻土助濾劑%電感耦閤等離子體質譜%重金屬%內標%動態反應池
규조토조려제%전감우합등리자체질보%중금속%내표%동태반응지
Diatomite filter aid%Inductively coupled plasma mass spectrometry%Heavy metal%Internal standard elements%Dynamic reaction cell
建立了硅藻土助滤剂中多个重金属元素的分析方法。以氢氟酸、硝酸和磷酸为混合试剂采用微波消解法对硅藻土助滤剂进行密闭消解,应用电感耦合等离子体质谱(ICP‐MS)法对消解液中的Be ,Cr ,Ni , As ,Cd ,Sb ,Sn ,Tl ,Hg ,Pb等10个元素进行测定。研究了高硅基体存在时所产的质谱干扰,首先通过优化仪器工作参数以及待测元素同位素的选择,最大程度地避免了质谱分析过程中的多种干扰,然后选择动态反应(DRC)技术消除了其它质谱干扰,多元素(Li ,Sc ,Y ,In ,Bi)内标溶液的使用消除了基体效应,维持了分析信号的稳定。分析结果显示,10个待测元素的检出限在3.29~15.68 ng?L -1之间,相对标准偏差(RSD )≤4.62%,加标回收率为90.71%~107.22%。方法的灵敏度高,准确度和精密度好,可用于硅藻土助滤剂的质量控制和安全评价。
建立瞭硅藻土助濾劑中多箇重金屬元素的分析方法。以氫氟痠、硝痠和燐痠為混閤試劑採用微波消解法對硅藻土助濾劑進行密閉消解,應用電感耦閤等離子體質譜(ICP‐MS)法對消解液中的Be ,Cr ,Ni , As ,Cd ,Sb ,Sn ,Tl ,Hg ,Pb等10箇元素進行測定。研究瞭高硅基體存在時所產的質譜榦擾,首先通過優化儀器工作參數以及待測元素同位素的選擇,最大程度地避免瞭質譜分析過程中的多種榦擾,然後選擇動態反應(DRC)技術消除瞭其它質譜榦擾,多元素(Li ,Sc ,Y ,In ,Bi)內標溶液的使用消除瞭基體效應,維持瞭分析信號的穩定。分析結果顯示,10箇待測元素的檢齣限在3.29~15.68 ng?L -1之間,相對標準偏差(RSD )≤4.62%,加標迴收率為90.71%~107.22%。方法的靈敏度高,準確度和精密度好,可用于硅藻土助濾劑的質量控製和安全評價。
건립료규조토조려제중다개중금속원소적분석방법。이경불산、초산화린산위혼합시제채용미파소해법대규조토조려제진행밀폐소해,응용전감우합등리자체질보(ICP‐MS)법대소해액중적Be ,Cr ,Ni , As ,Cd ,Sb ,Sn ,Tl ,Hg ,Pb등10개원소진행측정。연구료고규기체존재시소산적질보간우,수선통과우화의기공작삼수이급대측원소동위소적선택,최대정도지피면료질보분석과정중적다충간우,연후선택동태반응(DRC)기술소제료기타질보간우,다원소(Li ,Sc ,Y ,In ,Bi)내표용액적사용소제료기체효응,유지료분석신호적은정。분석결과현시,10개대측원소적검출한재3.29~15.68 ng?L -1지간,상대표준편차(RSD )≤4.62%,가표회수솔위90.71%~107.22%。방법적령민도고,준학도화정밀도호,가용우규조토조려제적질량공제화안전평개。
This study established a method for determining Be ,Cr ,Ni ,As ,Cd ,Sb ,Sn ,Tl ,Hg and Pb ,total 10 heavy metals in diatomite filter aid .The diatomite filter aid was digested by using the mixture acid of HNO3 + HF+ H3 PO4 in microwave sys‐tem ,10 heavy metals elements were determined by inductively coupled plasma mass spectrometry (ICP‐MS) .The interferences of mass spectrometry caused by the high silicon substrate were optimized ,first the equipment parameters and isotopes of test metals were selected to eliminate these interferences ,the methane was selected as reactant gas ,and the mass spectral interfer‐ences were eliminated by dynamic reaction cell (DRC) .Li ,Sc ,Y ,In and Bi were selected as the internal standard elements to correct the interferences caused by matrix and the drift of sensitivity .The results show that the detection limits for analyte is in the range of 3.29~15.68 ng?L -1 ,relative standard deviations (RSD) is less than 4.62% ,and the recovery is in the range of 90.71% ~107.22% .The current method has some advantages such as ,high sensitivity ,accurate ,and precision ,which can be used in diatomite filter aid quality control and safety estimations .