分光光度法%氯乙酸%影响
分光光度法%氯乙痠%影響
분광광도법%록을산%영향
spectrophotometric%chloroacetic acid%influence
主要对分光光度法测定水样中三氯乙酸实验进行了研究,并着重分析了溶液中的二氯乙酸,一氯乙酸对三氯乙酸的测定实验结果的影响。结果表明,当水样中只有三氯乙酸时,在530 nm处进行吸光度法检测,其线性检测范围为0~10 mg/L,相关线斜率为0.1106,相关系数R2=0.9999,最大加标回收率为99.98%,最小相对标准偏差为1.78%。当三氯乙酸水样中有等浓度的二氯乙酸时,相关线斜率为0.1087,且对等浓度二氯乙酸进行530 nm处分光光度法检测时,二氯乙酸的最大影响百分比达8.73%;当三氯乙酸水样中有等浓度的一氯乙酸时,相关线斜率为0.1118,且对等浓度一氯乙酸进行530 nm处分光光度法检测时,一氯乙酸的最大影响百分比最大达3.49%。
主要對分光光度法測定水樣中三氯乙痠實驗進行瞭研究,併著重分析瞭溶液中的二氯乙痠,一氯乙痠對三氯乙痠的測定實驗結果的影響。結果錶明,噹水樣中隻有三氯乙痠時,在530 nm處進行吸光度法檢測,其線性檢測範圍為0~10 mg/L,相關線斜率為0.1106,相關繫數R2=0.9999,最大加標迴收率為99.98%,最小相對標準偏差為1.78%。噹三氯乙痠水樣中有等濃度的二氯乙痠時,相關線斜率為0.1087,且對等濃度二氯乙痠進行530 nm處分光光度法檢測時,二氯乙痠的最大影響百分比達8.73%;噹三氯乙痠水樣中有等濃度的一氯乙痠時,相關線斜率為0.1118,且對等濃度一氯乙痠進行530 nm處分光光度法檢測時,一氯乙痠的最大影響百分比最大達3.49%。
주요대분광광도법측정수양중삼록을산실험진행료연구,병착중분석료용액중적이록을산,일록을산대삼록을산적측정실험결과적영향。결과표명,당수양중지유삼록을산시,재530 nm처진행흡광도법검측,기선성검측범위위0~10 mg/L,상관선사솔위0.1106,상관계수R2=0.9999,최대가표회수솔위99.98%,최소상대표준편차위1.78%。당삼록을산수양중유등농도적이록을산시,상관선사솔위0.1087,차대등농도이록을산진행530 nm처분광광도법검측시,이록을산적최대영향백분비체8.73%;당삼록을산수양중유등농도적일록을산시,상관선사솔위0.1118,차대등농도일록을산진행530 nm처분광광도법검측시,일록을산적최대영향백분비최대체3.49%。
Spectrophotometry of trichloroacetic acid were discussed in this paper,and the influence of dichloro-acetic acid and chloroacetic acidsolution in sulution of experimental results was emphatically analyzesed.It showed that with the absorbance detection at 530 nm and the linear detection range for 0mg/L-10mg/L,the slope of the line is 0.1106,the correlation coefficient is 0.9999,the maximum recovery rate is 99.98%,and the minimum relative standard deviation is 1.78%when there was only trichloroacetic acid in water.When there was the dichloroacetic acid with equal concertration of the trichloroacetic acid in the solution,the slope of the line is 0.1087 at 530 nm spectrophotometrically detection,and the greatest impact was 8.73%.When there was the chloroacetic acid with equal concertration of the trichloroacetic acid in the solution,the slope of the line is 0.1118 at 530 nm spectrophotometrically detection,and the greatest impact was 3.49%.