目的:检测口腔综合治疗机所存在的电气安全隐患,提高其安全性能和医护人员、患者的安全系数。方法采用奥利克288型电气安全检测仪,按照检测规范要求对我院综合科、牙体科、修复科以及儿童齿科4个科室共计100张椅位进行电气安全检测,检测项目主要包括外壳漏电流、患者漏电流、对地绝缘阻抗等6项指标。结果最终检测结果显示,不合格的指标主要是电源线接头氧化,以及治疗机地箱内湿度过大引起的。结论通过检测使我们意识到口腔综合治疗机在其特殊的环境下也存在不安全因素,应该对其各项指标进行定期检测,并纳入常态化管理,这也是医院设备质量控制体系中的重要环节。
目的:檢測口腔綜閤治療機所存在的電氣安全隱患,提高其安全性能和醫護人員、患者的安全繫數。方法採用奧利剋288型電氣安全檢測儀,按照檢測規範要求對我院綜閤科、牙體科、脩複科以及兒童齒科4箇科室共計100張椅位進行電氣安全檢測,檢測項目主要包括外殼漏電流、患者漏電流、對地絕緣阻抗等6項指標。結果最終檢測結果顯示,不閤格的指標主要是電源線接頭氧化,以及治療機地箱內濕度過大引起的。結論通過檢測使我們意識到口腔綜閤治療機在其特殊的環境下也存在不安全因素,應該對其各項指標進行定期檢測,併納入常態化管理,這也是醫院設備質量控製體繫中的重要環節。
목적:검측구강종합치료궤소존재적전기안전은환,제고기안전성능화의호인원、환자적안전계수。방법채용오리극288형전기안전검측의,안조검측규범요구대아원종합과、아체과、수복과이급인동치과4개과실공계100장의위진행전기안전검측,검측항목주요포괄외각루전류、환자루전류、대지절연조항등6항지표。결과최종검측결과현시,불합격적지표주요시전원선접두양화,이급치료궤지상내습도과대인기적。결론통과검측사아문의식도구강종합치료궤재기특수적배경하야존재불안전인소,응해대기각항지표진행정기검측,병납입상태화관리,저야시의원설비질량공제체계중적중요배절。
Objective To inspect the security problems for the oral comprehensive treatment machine so as to enhance its safety performance as well as the safety coefifcient for medical personnel and patients. Methods With application of Ehrlich 288 type electrical safety testing instrument, the electrical safety testing was performed for 100 chair positions according to the examination standard requirement of 4 departments in the hospital, including General Department, Department of Endodontics, Department of Prosthodontics and Department of Children’ Dental. The testing project mainly included 6 indexes such as casing leakage current, patient leakage current and the insulation impedance.Results The ifnal test results showed that the main index of the non-conformity was the oxidation of the power line and the over humidity in the treatment machine.Conclusion Through the testing, it had been realized that unsafe factors also existed in the oral treatment machine in special environment. Hence, regular detection of its indicators should be done and incorporated into the normal management, which also served as an important part in the hospital equipment quality control system.