Chinese Medical Record
2015年
11期
84-87
,共4页
王艳平%吕锐%王海阔%周剑%克德娜
王豔平%呂銳%王海闊%週劍%剋德娜
왕염평%려예%왕해활%주검%극덕나
乳腺X线摄影术%乳腺癌%辐射剂量%低剂量
乳腺X線攝影術%乳腺癌%輻射劑量%低劑量
유선X선섭영술%유선암%복사제량%저제량
Mammography%Breast cancer%Radiation dosage%Low-dose
目的:探讨低剂量摄影模式在乳腺X线检查中的价值。方法利用钼铑双靶乳腺X线机,采用标准模式(standard model,STD)和剂量优先模式(dose model,DOSE)对200例女性受检者双侧乳腺进行头尾位(cranial and caudal,CC)和内外斜位(mediolateral oblique,MLO)摄影,双盲法评价800幅乳腺图像质量,分别记录每名受检者乳腺平均腺体剂量值(average glandular dose,AGD)和皮肤入射剂量值(entrance skin exposure,ESE),对两种摄影模式的钙化显示率、病变阳性率和乳腺癌检出率进行对比分析。结果两种模式下乳腺图像均能够清晰显示正常腺体和病变结构;与STD模式相比,DOSE模式下受检者乳腺AGD降低17.21%,ESE降低12.78%,两者间均具有显著性差异,FAGD=44.070,P=0.000,FESE =14.896,P=0.000。DOSE模式下钙化显示率为46%,STD模式下钙化显示率为57%;两种模式下病变阳性率分别为14%和22%,两者间无显著性差异,x2=0.153,P=0.086;两种模式乳腺癌检出率分别为6%和7%,两者间无显著性差异,x2=0.752,P=0.605。结论在获取同等乳腺图像质量情况下,钼铑双靶乳腺X线检查中DOSE模式能够有效降低受检者的辐射剂量。
目的:探討低劑量攝影模式在乳腺X線檢查中的價值。方法利用鉬銠雙靶乳腺X線機,採用標準模式(standard model,STD)和劑量優先模式(dose model,DOSE)對200例女性受檢者雙側乳腺進行頭尾位(cranial and caudal,CC)和內外斜位(mediolateral oblique,MLO)攝影,雙盲法評價800幅乳腺圖像質量,分彆記錄每名受檢者乳腺平均腺體劑量值(average glandular dose,AGD)和皮膚入射劑量值(entrance skin exposure,ESE),對兩種攝影模式的鈣化顯示率、病變暘性率和乳腺癌檢齣率進行對比分析。結果兩種模式下乳腺圖像均能夠清晰顯示正常腺體和病變結構;與STD模式相比,DOSE模式下受檢者乳腺AGD降低17.21%,ESE降低12.78%,兩者間均具有顯著性差異,FAGD=44.070,P=0.000,FESE =14.896,P=0.000。DOSE模式下鈣化顯示率為46%,STD模式下鈣化顯示率為57%;兩種模式下病變暘性率分彆為14%和22%,兩者間無顯著性差異,x2=0.153,P=0.086;兩種模式乳腺癌檢齣率分彆為6%和7%,兩者間無顯著性差異,x2=0.752,P=0.605。結論在穫取同等乳腺圖像質量情況下,鉬銠雙靶乳腺X線檢查中DOSE模式能夠有效降低受檢者的輻射劑量。
목적:탐토저제량섭영모식재유선X선검사중적개치。방법이용목로쌍파유선X선궤,채용표준모식(standard model,STD)화제량우선모식(dose model,DOSE)대200례녀성수검자쌍측유선진행두미위(cranial and caudal,CC)화내외사위(mediolateral oblique,MLO)섭영,쌍맹법평개800폭유선도상질량,분별기록매명수검자유선평균선체제량치(average glandular dose,AGD)화피부입사제량치(entrance skin exposure,ESE),대량충섭영모식적개화현시솔、병변양성솔화유선암검출솔진행대비분석。결과량충모식하유선도상균능구청석현시정상선체화병변결구;여STD모식상비,DOSE모식하수검자유선AGD강저17.21%,ESE강저12.78%,량자간균구유현저성차이,FAGD=44.070,P=0.000,FESE =14.896,P=0.000。DOSE모식하개화현시솔위46%,STD모식하개화현시솔위57%;량충모식하병변양성솔분별위14%화22%,량자간무현저성차이,x2=0.153,P=0.086;량충모식유선암검출솔분별위6%화7%,량자간무현저성차이,x2=0.752,P=0.605。결론재획취동등유선도상질량정황하,목로쌍파유선X선검사중DOSE모식능구유효강저수검자적복사제량。
Objective To investigate the value of the low-dose mode of molybdenum-rhodium target X-ray photography in mammary examination.Methods Two hundred female clients were recruited,and their bilateral mammographic images were prospectively acquired by the standard mode(STD) and dose mode(DOSE) in both cranial and caudal(CC) position and mediolateral oblique(MLO) position using molybdenum-rhodium target X-ray machine.The total of 800 mammograms were obtained and analyzed to assess the image quality with double-blind method.The value of average glandular dose(AGD) and entrance skin exposure(ESE) on each client were detected and recorded respectively.The contrast analysis between STD and DOSE on display rate of calcification,positive rate of lesion and detection rate of breast cancer were performed.Results The mammary gland and characteristics of lesion were clearly displayed in both STD and DOSE images.Compared with the STD,the reduction of X-ray exposure such as AGD and ESE values detected by DOSE was 17.21% and 12.78% respectively, there was a significant difference in AGD and ESE values between the two modes.FAGD = 44.070,P = 0.000, FESE=14.896,P= 0.000.The display rate of calcification was 46% in DOSE and 57% in STD.The positive rate of lesion was 14% and 22% in DOSE and STD respectively,there was no significant difference in the positive rate of lesion between DOSE and STD,x2= 0.153,P = 0.086.The detection rate of breast cancer was 6% and 7% in two modes respectively,The ANOVA results showed that the differences of the detection rate of breast cancer between STD and DOSE was no significant,x2= 0.752,P = 0.605.Conclusions Under the condition of achieving equivalent quality of mammograms,the dose mode may effectively decrease radiation exposure on client in molybdenum- rhodium target mammography.