中南大学学报(英文版)
中南大學學報(英文版)
중남대학학보(영문판)
Journal of Central South University
2015年
12期
4566-4577
,共12页
李坤%潘博%高文朋%封海波%付宜利%王树国
李坤%潘博%高文朋%封海波%付宜利%王樹國
리곤%반박%고문붕%봉해파%부의리%왕수국
force feedback%force/torque sensor%Stewart platform%optimal design%robot-assisted minimally invasive surgery
In order to restore force sensation to robot-assisted minimally invasive surgery (RMIS), design and performance evaluation of a miniature 6-axis force/torque sensor for force feedback is presented. Based on the resistive sensing method, a flexural-hinged Stewart platform is designed as the flexible structure, and a straightforward optimization method considering the force and sensitivity isotropy of the sensor is proposed to determine geometric parameters which are best suited for the given external loads. The accuracy of this method is preliminarily discussed by finite element methods (FEMs). The sensor prototype is fabricated with the development of the electronic system. Calibration and dynamic loading tests for this sensor prototype are carried out. The working ranges of this sensor prototype are 30 N and 300 N·mm, and resolutions are 0.08 N in radial directions, 0.25 N in axial direction, and 2.4 N·mm in rotational directions. It also exhibits a good capability for a typical dynamic force sensing at a frequency close to the normal heart rate of an adult. The sensor is compatible with surgical instruments for force feedback in RMIS.