目的 对儿童散瞳后应用Lenstar和IOLMaster测量眼球生物结构参数,并对2种仪器测量结果进行再现性和一致性评价.方法 前瞻性研究.75例(75眼)应用1%盐酸环喷托酯滴眼液快速散瞳,2位操作者分别应用Lenstar和IOLMaster测量其眼轴长度(AL)、最小角膜屈光力径线上角膜曲率(Kf)、最大角膜屈光力径线上角膜曲率(Ks)、平均角膜曲率(Km)、角膜水平直径(WTW)、前房深度(ACD).Lenstar的再现性通过组内标准差(Sw)、重测系数(TRT)、组内变异系数(CV)和组内相关系数(ICC)进行评估,采用配对t检验评价2种仪器差异性,Bland-Altman法分析2种仪器的一致性.结果 Lenstar测得的AL、Kf、Ks、Km和ACD具有较好的再现性,其TRT分别为0.03 mm、0.33 D、0.22 D、0.03 D和0.23 mm,所有ICC均>0.99.Lenstar和IOLMaster测量ACD、WTW、Ks和Km分别为(3.77±0.20)mm和(3.66±0.20)mm(t=16.194,P<0.01),(12.11±0.43)mm和(12.41±0.53)mm (t=-5.973, P<0.01),(43.62±1.49)D和(43.66±1.49)D(t=-2.277,P<0.05),(43.08±1.36)D和(43.12±1.35)D(t=-2.673,P<0.01),差异均有统计学意义.2种仪器测量ACD、AL、Kf、Ks、Km的95%一致性区间(LoA)较小,其上下限最大绝对值分别为0.23 mm、0.07 mm、0.29 D、0.35 D和0.29 D,一致性较好,而WTW的95% LoA较大(-1.17~0.57 mm),一致性较差.结论 Lenstar在不同操作者中再现性良好;Lenstar与IOLMaster测量散瞳后儿童的WTW有差别,一致性较差,但是Lenstar更有优势;测量ACD、AL、Kf、Ks以及Km一致性较好,可以相互参考使用.
目的 對兒童散瞳後應用Lenstar和IOLMaster測量眼毬生物結構參數,併對2種儀器測量結果進行再現性和一緻性評價.方法 前瞻性研究.75例(75眼)應用1%鹽痠環噴託酯滴眼液快速散瞳,2位操作者分彆應用Lenstar和IOLMaster測量其眼軸長度(AL)、最小角膜屈光力徑線上角膜麯率(Kf)、最大角膜屈光力徑線上角膜麯率(Ks)、平均角膜麯率(Km)、角膜水平直徑(WTW)、前房深度(ACD).Lenstar的再現性通過組內標準差(Sw)、重測繫數(TRT)、組內變異繫數(CV)和組內相關繫數(ICC)進行評估,採用配對t檢驗評價2種儀器差異性,Bland-Altman法分析2種儀器的一緻性.結果 Lenstar測得的AL、Kf、Ks、Km和ACD具有較好的再現性,其TRT分彆為0.03 mm、0.33 D、0.22 D、0.03 D和0.23 mm,所有ICC均>0.99.Lenstar和IOLMaster測量ACD、WTW、Ks和Km分彆為(3.77±0.20)mm和(3.66±0.20)mm(t=16.194,P<0.01),(12.11±0.43)mm和(12.41±0.53)mm (t=-5.973, P<0.01),(43.62±1.49)D和(43.66±1.49)D(t=-2.277,P<0.05),(43.08±1.36)D和(43.12±1.35)D(t=-2.673,P<0.01),差異均有統計學意義.2種儀器測量ACD、AL、Kf、Ks、Km的95%一緻性區間(LoA)較小,其上下限最大絕對值分彆為0.23 mm、0.07 mm、0.29 D、0.35 D和0.29 D,一緻性較好,而WTW的95% LoA較大(-1.17~0.57 mm),一緻性較差.結論 Lenstar在不同操作者中再現性良好;Lenstar與IOLMaster測量散瞳後兒童的WTW有差彆,一緻性較差,但是Lenstar更有優勢;測量ACD、AL、Kf、Ks以及Km一緻性較好,可以相互參攷使用.
목적 대인동산동후응용Lenstar화IOLMaster측량안구생물결구삼수,병대2충의기측량결과진행재현성화일치성평개.방법 전첨성연구.75례(75안)응용1%염산배분탁지적안액쾌속산동,2위조작자분별응용Lenstar화IOLMaster측량기안축장도(AL)、최소각막굴광력경선상각막곡솔(Kf)、최대각막굴광력경선상각막곡솔(Ks)、평균각막곡솔(Km)、각막수평직경(WTW)、전방심도(ACD).Lenstar적재현성통과조내표준차(Sw)、중측계수(TRT)、조내변이계수(CV)화조내상관계수(ICC)진행평고,채용배대t검험평개2충의기차이성,Bland-Altman법분석2충의기적일치성.결과 Lenstar측득적AL、Kf、Ks、Km화ACD구유교호적재현성,기TRT분별위0.03 mm、0.33 D、0.22 D、0.03 D화0.23 mm,소유ICC균>0.99.Lenstar화IOLMaster측량ACD、WTW、Ks화Km분별위(3.77±0.20)mm화(3.66±0.20)mm(t=16.194,P<0.01),(12.11±0.43)mm화(12.41±0.53)mm (t=-5.973, P<0.01),(43.62±1.49)D화(43.66±1.49)D(t=-2.277,P<0.05),(43.08±1.36)D화(43.12±1.35)D(t=-2.673,P<0.01),차이균유통계학의의.2충의기측량ACD、AL、Kf、Ks、Km적95%일치성구간(LoA)교소,기상하한최대절대치분별위0.23 mm、0.07 mm、0.29 D、0.35 D화0.29 D,일치성교호,이WTW적95% LoA교대(-1.17~0.57 mm),일치성교차.결론 Lenstar재불동조작자중재현성량호;Lenstar여IOLMaster측량산동후인동적WTW유차별,일치성교차,단시Lenstar경유우세;측량ACD、AL、Kf、Ks이급Km일치성교호,가이상호삼고사용.
Objective To investigate the reproducibility and agreement between Lenstar and IOLMaster when measuring axial length (AL), anterior chamber depth (ACD), keratometry for the flattest meridian (Kf), keratometry for the steepest meridian (Ks), mean keratometry (Km) and corneal diameter (WTW) of children after dilation.Methods In this prospective clinical study,75 eligible children were recruited.One eye of each subject was randomly selected and all subjects were dilated with 1% cyclopentolate.All subjects were examined by two examiners for all parameters (AL, Kf, Ks, Km, WTW, ACD).Within-subject standard deviation (Sw), test-retest repeatability (TRT), the within-subject coefficient of variation (CV) and the interclass correlation coefficient (ICC) were used to analyze reproducibility.The differences between these ocular parameters were analyzed using a paired t-test.Bland-Altman was used to assess the agreement of the instruments.Results The reproducibility results of AL, Kf, Ks, Km and ACD measured by the two examiners using Lenstar were good.The TRT values were 0.03 mm, 0.33 D, 0.22 D, 0.03 D and 0.23 mm,respectively.All the ICC values were over 0.99.The mean ACD, WTW, Ks and Km obtained by Lenstar and IOLMaster were 3.77±0.20 mm and 3.66±0.20 mm (t=16.194, P<0.01), 12.11±0.43 mmand 12.41±0.53 mm (t=-5.973, P<0.01), 43.62±1.49 D and 43.66±1.49 D (t=-2.277, P<0.05), and 43.08±1.36 D and 43.12±1.35 D (t=-2.673, P<0.O1), respectively.All these values showed significant differences between the two methods.Bland-Altman analysis showed that the two instruments had comparable results for ACD, AL, Kf, Ks, and Km.The maximum absolute values of the 95% limits of agreement (LoA) were 0.23 mm, 0.07 mm, 0.29 D, 0.35 D and 0.29 D.But the 95% LoA ofWTW was large (-1.17~0.57 mm), indicating the agreement of WTW between these two instruments was poor.Conclusion The reproducibility is good for Lenstar.The WTW difference between Lenstar and IOLMaster is statistically significant, and the consistency is poor, but Lenstar is much better.Otherwise, the consistency of ACD, AL, Kf, Ks and Km between Lenstar and IOLMaster is good and could be regarded as clinically interchangeable.